adjusted linear actuator, replaced Nylon 0.375" OD then Calibrated. Found Servo error more than 1 % and Tube 0.18"OD were modified to 0.25" copper tube. Agr will send quotation for tubing set again.
adjusted linear actuator, replaced Nylon 0.375" OD then Calibrated. Found Servo error more than 1 % and Tube 0.18"OD were modified to 0.25" copper tube. Agr will send quotation for tubing set again.actuator