3.1. Analytical quality assurance
The analytical characteristics of the methods used to measure
Se, Cu, Cr, and Mn were evaluated, establishing the detection limits
(Table 1). The limits of detection (LOD) and quantification (LOQ)
were calculated with three and ten times the standard deviation
of the blank signal divided by the slope of the analytical curve (sensitivity),
respectively