PAR concentration was tested in the range 0.005± 0.1%. Results obtained show that signals became constants for PAR concentration over 0.01%. So a 0.05% PAR concentration was selected. Effect of pH on the nickel determination with PAR reagentwas also studied. Results showed that the absorbance signal is maximum about pH 10. So, a borate buffer solution was employed, according to previous study [24].