Test Cards
4.1 TanDelta - PF Test Card
4.1.1 Main Page (1/2)
Select "Assessment" to automatically assess the test, clear for no assessment.
Enter the nominal values in the entry fields (here "C ref" and "DF ref"; availability and naming depend on the measuring mode). These values serve as reference for the assessment. Their tolerance range can be set on the Settings Page (refer to 4.1.3 on page 60).
A measurement is rated as ’OK’ if both values are within their tolerance range. The assessment is displayed in the test point tables’s column "?"
Note: While a test is running, new nominal values can already be entered.