Method comparison data were collected using CLSI guideline EP9-A4. Venous blood samples were
collected in lithium heparin evacuated tubes from hospital patients. The whole blood samples were run
in duplicate on the i-STAT System. The samples were then centrifuged to obtain plasma and analyzed in
duplicate on the comparative instrument. All samples were analyzed on both methods within 15 minutes
of each other