Instrumentation
Analysis were carried out with an inductively coupled plasma
mass spectrometer equipped with a reaction cell (DRC-ICP-MS
ELAN DRCII, Perkin Elmer, SCIEX, Norwalk, CT, USA) operating
with high-purity argon (99.999%, Praxaair, Brazil). Sample
introduction system included a quartz cyclonic spray chamber
and a Meinhards nebulizer connected by Tygons tubes to the
ICP-MS’s peristaltic pump (set at 20 rpm). The ICP-MS was
operated with Pt sampler and skimmer cones purchased either
from Perkin Elmer. A radiofrequency (rf) of 1100 W power was
selected in pulse mode with autolens one. Sample data were
acquired by using 20 sweeps/reading, 1 reading/replicate and a
50 ms dwell time. Argon nebulizer gas flow rate was optimized
daily from 0.5 to 0.9 L min-
1
. Data were acquired in counts
per second (cps). The 208Pb isotope was selected.
เครื่องมือวัดการวิเคราะห์ได้ดำเนินการด้วยการอุปนัยคู่พลาสมาแมสสเปกโทรมิเตอร์พร้อมกับปฏิกิริยาเซลล์ ( drc-icp-msรวม drcii เพอร์กินเอลเมอร์ sciex , , , Norwalk , CT , สหรัฐอเมริกา ) ปฏิบัติการที่มีความบริสุทธิ์สูงอาร์กอน ( 99.999 % praxaair , บราซิล ) ตัวอย่างระบบแนะนำรวมควอทซ์ใช้สเปรย์ในห้องและ meinhards nebulizer เชื่อมต่อโดย tygons หลอดไปICP-MS ของ peristaltic ปั๊ม ( ตั้งไว้ที่ 20 รอบต่อนาที ) ส่วน ICP-MS คือดำเนินการใน PT และ Skimmer กรวยซื้ออย่างใดอย่างหนึ่งจากเพอร์กินเอลเมอร์ เป็นด้วยคลื่นวิทยุ ( RF ) 1100 W มีอำนาจเลือกในโหมดชีพจรกับ autolens 1 ข้อมูล ตัวอย่างคือข้อมูลที่ได้จากการใช้ 20 กวาด / อ่าน 1 / อ่านซ้ำและ50 นางสาวเวลาหน่วง อัตราการไหลของก๊าซอาร์กอน nebulizer ก็เหมาะทุกวันตั้งแต่ 0.5 ถึง 0.9 ลิตร มิน1. ข้อมูลที่ได้มา นับต่อวินาที ( CPS ) 208pb ไอโซโทปที่ถูกเลือก