2.2. HPLC-MS conditions
HPLC analyses were carried out using an Agilent 1200 series HPLC system (Agilent, Waldbronn, Germany). The system com-prised a vacuum degasser, binary pumps, a thermostated autosam-pler and column compartment. The column used for the chromatography was a Prontosil C-18 (Bischoff, 3 150 mm, 3 lm) at a flow rate of 0.5 ml/min and a column temperature of 35 C. Injection volume was 10 ll. The HPLC mobile phases con-sisted of 0.1% aqueous formic acid (A) and acetonitrile (B). Follow-ing linear gradient was used (%B): 0–5 min (5%), 15 min (28%), and 16–20 min (100%). The column was then re-equilibrated for 10 min, making a total run time of 30 min. A switching valve was used in order to protect the instrument from unnecessary matrix pollution. The HPLC eluent flow between 12 and 18 min was direc-ted into the mass spectrometer while the rest of the run went to waste. The FA signal was detected at 15 min.
2.2 . hplc-ms เงื่อนไข
HPLC วิเคราะห์ทดลองใช้เลนต์ 1200 ชุด 2 ระบบ ( Agilent , waldbronn , เยอรมัน ) prised com ระบบสุญญากาศ degasser , ปั๊มไบนารี , thermostated โต้แซมก็ช่องในคอลัมน์ คอลัมน์ที่ใช้สำหรับโครมาโตกราฟี คือ พรอนโตซิล ( บิชอป - 3 , 150 มม. 3 อิม ) ที่flโอ๊ยอัตรา 0.5 มิลลิลิตรต่อนาทีและอุณหภูมิ 35 องศาเซลเซียส คอลัมน์ปริมาณการฉีดได้ 10 จะ พบมือถือเฟสคอน sisted 0.1% กรดฟอร์มิกสารละลายไน ( A ) และ ( B ) ไอเอ็นจีตามลาดเชิงใช้ ( % B ) : 0 – 5 นาที ( 5% ) , 15 นาที ( 28% ) และ 16 – 20 นาที ( 100% ) คอลัมน์เป็นแล้ว equilibrated 10 นาที ทำให้เวลาวิ่ง 30 นาที เปลี่ยนวาล์วถูกใช้เพื่อปกป้องจากมลภาวะในเมทริกซ์ที่ไม่จำเป็นพบตัวflโอ๊ยระหว่าง 12 และ 18 นาทีตรง เท็ดเป็นแมสสเปกโทรมิเตอร์ในขณะที่ส่วนที่เหลือของรันไปเสีย สัญญาณเอฟเอ ที่ตรวจพบใน 15 นาที