If the Aux Terminal Mode is configured for Auto Calibration, be sure the jumper remains across the auxiliary terminals until the
Field Communicator prompts you to remove it. Removing the jumper too soon will cause the instrument to begin auto calibration.
Auto calibration can be terminated by shorting the Aux Terminals for one second.