INSTRUCTION MANUAL
Laboratory Equipment Pty Ltd. 2 LEVEL INCUBATOR SHAKER 2013 VER1.01
LABORATORY EQUIPMENT
PTY LTD
WWW.LABEC.COM.AU
TWO LEVEL
INCUBATOR SHAKER
USER’S MANUAL
Content
1. Safety Instructions
2. Description
3. Control Panel
4. Starting the Unit
5. Temperature Control
6. Overtemperature Alarm
7. Speed Control
8. Timer Function
9. Defrosting
10. Non-volatile Memory
11. Technical Specifications
12. Trouble-shooting
13. Packing List
Laboratory Equipment Pty Ltd. 2 LEVEL INCUBATOR SHAKER 2013 VER1.01
Thank you very much for choosing ZHWY-2102/2112 series Two Layers Incubator
Shakers. To ensure a smooth operation of this unit, we would like to ask you to read
this instruction manual thoroughly, and always operate the machine as guided by this
manual. Also keep this manual with the unit for maintenance reference. In case you
meet problems with this unit, please feel free to contact Laboratory Equipment Pty
Ltd. We will help you identify the problem and provide trouble-shooting service for
you.
1. Safety Instructions
1. The unit must be properly wired to ensure safety either to the operator or facilities. It is
recommended to use dedicated line to avoid interference from other equipment.
2. Please verify the local electric power supply is in accordance with the electrical requirement
240V 50/60Hz of the unit.
3. Do not use extra power cords or lines for this unit, do not bend, entangle or twist the power
cord, or apply excessive tension to it, otherwise it may cause damage to the unit or result in
electrical shock or fire.
4. Do not touch the unit with wet hands or insert other items into the wind inlet when the unit is
running. It may cause a fire or electrical shock.
5. Do not move the unit when it is operating.
6. In case moving the unit is necessary, please be sure the unit is kept upright. Otherwise, it may
cause damage to the refrigeration system inside.
7. It is good practice to ensure that the shaking platform comes to a complete standstill before
open the door of the chamber.
8. When opening or closing the chamber door, please handle with care, do not use too much
force.
9. In case fuse has to be replaced, the electric power supply has to be cut off.
10. The unit must be placed on solid, flat floors near a grounded electrical outlet, allow sufficient
clearance on all sides of the unit, at least 20cm, for proper ventilation. Avoid locations under
direct sunlight or near heat producing equipments such as an oven.
11. After continuous refrigeration for 10 days, it is good practice to de-damp the unit by turning
on the heating function.
12. Regular check to the flask clamps is necessary to ensure they are in good status.
13. Frequent opening of the chamber door is not recommended as this may affect the temperature
recovery.
14. The unit must be kept away from volatile, flammable, explosive liquids or gases. This may
cause damages either to the unit or property losses includes facilities and buildings.
15. In case of malfunction or burning smell, the unit must be immediately ceased for checking.
16. If a thunderstorm approaches in the distance, immediately turn off the power. Lightning from
the storm may cause fire or electrical shock.
17. The above safety instructions must be followed during the operation of the unit. All persons
operating this unit should review this manual. No compensation or claim can be made against
the manufacturer if damage is caused to the personnel or facilities that result from the
improper use of this unit.
Laboratory Equipment Pty Ltd. 2 LEVEL INCUBATOR SHAKER 2013 VER1.01
2. Unit Description
ZHWY-2102/2112 series Two Layers Incubator Shakers are designed for large capacity growth
of biological organisms under optimum conditions of temperature, time and agitation. These
microprocessor controlled
incubator shakers offer a number
of advanced features include
user-friendly digital setpoints
keypad, overtemperature alarm
and self-correcting feedback
controls on temperature and
shaking speed.
Features in Summary
l Powerful triple eccentric
drive mechanism guarantees
a lifetime of smooth and
continuous shaking motion.
l A P.I.D microprocessor
controller maintains precise
temperature and shaking
speed.
l User-friendly sealed keypad
allows easy digital setting
on running time, speed, and
optimum temperature
condition.
l Large, easy-to-read LCD
display presents a menu of
functional readouts.
l Automatic shutoff at end of
study via electronic timer, adjustable from 1 minute to 500 hours.
l Automatic shutoff of shaker when lid is opened which ensures operator safety.
l Non-volatile memory retains setpoints during a power failure and restarts shaker
automatically.
l Alarms programmed to warn of setpoints deviation and to indicate when timer operation has
expired.
l A high volume air circulation system maintains optimum temperature uniformity, as well as
providing maximum oxygen transfer.
l See-through chamber door provides complete visibility and minimize the need to open the
chamber frequently.
l Mirror-finished stainless steel interior provides excellent durability and an easy-to-clean
surface.
Laboratory Equipment Pty Ltd. 2 LEVEL INCUBATOR SHAKER 2013 VER1.01
3. Control Panel
4. Starting the Unit
Turn on the power to the unit by pressing the Main Switch located on the right side
of the unit body, the unit enters into the running mode. All keys on the control panel
are activated in the running mode. The LED Display shows the preset temperature,
speed, timer and finally, measured temperature. Press the Start/Pause key, the unit
will operate on the preset parameters. Second press the Start/Pause key while the unit
is running stops the operation. Holding the Power On/Off key for more than 3 seconds,
the unit is powered off.
5. Temperature Control
Temperature of the unit can be set from ambient-20°C to 60°C. Temperature setting is
completed by the following steps:
1. Press the Set/Confirmation key, then press the Increase key to enter the
password 3.
2. Press the Temperature key, the unit enters into temperature setting mode.
Temperature value can be set via the Increase/Decrease key. To confirm the
new setting, press the Set/Confirmation key.
Setting of a new value overwrites the old value in the memory.
Laboratory Equipment Pty Ltd. 2 LEVEL INCUBATOR SHAKER 2013 VER1.01
Refrigeration
Press the Set/Confirmation key, then press the Increase key to enter the password 9.
Press the Refrigeration key, the unit enters into refrigeration setting mode.
Refrigeration parameter can be set via the Increase/Decrease key. To confirm the new
setting, press the Set/Confirmation key.
• If the unit is to be used in the temperature range of ambient-5°C to ambient+5°C, the
refrigeration parameter can be set at the value 0.2.
• If the unit is to be used in the temperature range of ambient-20°C to ambient-5°C, the
refrigeration parameter can be set at the value 0.5.
• If the unit is to be used in the temperature range of ambient+5°C to 60°C, the
refrigeration parameter can be set at the value 2. The refrigeration is cancelled.
6. Overtemperature Alarm
In case the gap between measured temperature and setting temperature excels ± 3°C,
the unit sends out audible and visible alarm automatically, and the yellow alarm
indicator flashes. The platform keeps on shaking while the heating or refrigeration
stops. Press the Temperature key will cancel the alarm for 30 minutes. Reset the
temperature value, the unit exits the alarm mode.
7. Speed Control
To set the desired shaking speed, first press the Set/Confirmation key, then press the
Increase key to enter the password 3. Press the Speed key, the unit enters into speed
setting mode. The desired speed value can be set via the Increase/Decrease key. To
confirm the new setting, press the Set/Confirmation key. Setting of a new value
overwrites the old value in the memory.
8. Timer
When timer function is needed, first press the Set/Confirmation key, then press the
Increase key to enter the password 3. Press the Timer key, the unit enters into timer
setting mode, the desired timer value can be set via the Increase/Decrease key. To
confirm the new setting, press the Set/Confirmation key.
By the time the remaining time reaches 0:00, the platform stops shaking, the unit sends out audible
and visible alarm for 2 minutes. Press the Timer key, the alarm will be muted.
Laboratory Equipment Pty Ltd. 2 LEVEL INCUBATOR SHAKER 2013 VER1.01
9. Defrosting
Operation the unit at a low temperature for a long period of time causes the buildup of
frost, which degrades the freezing performance and may disable the temperature from
being controlled at the setpoint. The defrosting function guarantees the unit to operate
smoothly and continuously in such a case. Defrosting function is set via the following
steps:
• First press the Set/Confirmation key, then press the Increase key to enter the
password 9. Press the Refrigeration key, the unit enters into refrigeration mode
(first stage).
• Press the Refrigeration key again, the unit enters into defrosting timer setting
mode (second stage). Defrosting timer value can be set via the
Increase/Decrease key.
• Press the Refrigeration key once more, the unit enters into defrosting time
period setting mode. Defrosting time period can be set via the
Increase/Decrease key.
• To confirm the new setting, press the Set/Confirmation key.
Defrosting Timer Setting refers to the function by which the unit starts defrosting every given time
period. For example, if the defrosting timer value is set at 1, the unit starts defrosting every 1 hour.
If the defrost timer value is set at 2, the unit starts defrosting every 2 hour. However, if the defrost
timer value is set at 0, the unit does not start defrosting at all.
Defrosting Time Period refers to the time span from the moment the unit starts defrosting till it
exits d
0/5000
จาก: -
เป็น: -
ผลลัพธ์ (ไทย) 1: [สำเนา]
คัดลอก!
www.labec.com.auห้องปฏิบัติการอุปกรณ์ Pty Ltdอีเมล์: sales@labec.com.auPh: 02 9560 2811 •แฟ็กซ์: 02 9560 6131ปี 2014คู่มือห้องปฏิบัติการอุปกรณ์ Pty Ltd. 2 ระดับ INCUBATOR เชคเกอร์ 2013 VER1.01อุปกรณ์ห้องปฏิบัติการPTY LTDWWW.LABEC.COM.AUระดับที่สองเชคเกอร์บ่มเพาะวิสาหกิจคู่มือการใช้เนื้อหา1. แนะนำความปลอดภัย2. คำอธิบาย3. ควบคุม4. หน่วยเริ่มต้น5. อุณหภูมิควบคุม6. overtemperature ปลุก7. ความเร็วในการควบคุม8. ฟังก์ชันจับเวลา9. การละลายน้ำแข็ง10. หน่วยความจำไม่ใช่ระเหย11. เทคนิค12. trouble-shooting13. บันทึกรายการห้องปฏิบัติการอุปกรณ์ Pty Ltd. 2 ระดับ INCUBATOR เชคเกอร์ 2013 VER1.01ขอบคุณมากสำหรับการเลือกชุด ZHWY-2102/2112 บ่มเพาะวิสาหกิจชั้นสองสโมสร เพื่อให้การดำเนินงานที่ราบรื่นของหน่วยนี้ เราอยากจะขอให้คุณอ่านคู่มือนี้อย่างละเอียด และมักจะมีเครื่องแนะนำตามนี้ด้วยตนเอง ยัง เก็บคู่มือนี้ มีหน่วยสำหรับการอ้างอิงการบำรุงรักษา ในกรณีที่คุณพบปัญหากับหน่วยนี้ กรุณาติดต่อห้องปฏิบัติการอุปกรณ์ Ptyจำกัด เราจะช่วยให้คุณระบุปัญหา และให้บริการ trouble-shootingคุณ1. แนะนำความปลอดภัย1. หน่วยต้องเป็นสายเพื่อเพิ่มความปลอดภัยผู้ให้บริการหรือสิ่งอำนวยความสะดวกอย่างเหมาะสม จึงแนะนำให้ใช้บรรทัดเฉพาะเพื่อหลีกเลี่ยงสัญญาณรบกวนจากอุปกรณ์อื่น ๆ2. กรุณาตรวจสอบเครื่องไฟฟ้าไฟเป็นไปตามความต้องการไฟฟ้า240V 50 / 60Hz ของหน่วย3. ทำไม่ใช้เสริมพลังงานสายหรือบรรทัดในหน่วยนี้ ไม่งอ entangle หรือบิดไฟสายไฟ หรือใช้แรงมากเกินไปนั้น มิฉะนั้นมันอาจก่อให้เกิดความเสียหายกับหน่วย หรือผลไฟฟ้าช็อตหรือไฟไหม้4. สัมผัสหน่วยมือเปียก หรือแทรกรายการอื่นเป็นทางเข้าของลมเมื่อหน่วยทำงานอยู่ มันอาจทำให้เกิดไฟไหม้หรือไฟฟ้าช็อต5. ย้ายหน่วยเมื่อทำงาน6. ในกรณีย้ายหน่วยมีความจำเป็น โปรดให้แน่ใจว่า หน่วยอยู่ตรง มันอาจเป็นอย่างอื่นทำให้เกิดความเสียหายระบบเครื่องทำความเย็นภายใน7. เป็นการดีเพื่อให้แน่ใจว่า แพลตฟอร์มงก ๆ มาติดสมบูรณ์ก่อนเปิดประตูห้อง8. เมื่อเปิด หรือปิดประตูหอการค้า กรุณาจัดการกับดูแล ไม่ใช้มากเกินไปกองทัพ9. ในกรณีที่ฟิวส์ได้ถูกแทนที่ ไฟไฟฟ้าได้ถูกตัดออก10. หน่วยที่ต้องวางบนพื้นแข็ง แบนใกล้เต้าเสียบไฟฟ้าป่นเล็กน้อย ให้เพียงพอเคลียร์ทุกด้านของหน่วย น้อย 20 ซม. การระบายอากาศที่เหมาะสม หลีกเลี่ยงสถานภายใต้แสงแดดโดยตรง หรือใกล้ผลิตอุปกรณ์เช่นเตาอบความร้อน11. หลังจากแช่แข็งอย่างต่อเนื่อง 10 วัน ดีที่จะยกเลิกบันทอนหน่วย โดยเปิดในการทำงานความร้อน12. ปกติเช็ค clamps หนาวจำเป็นต้องอยู่ในสถานะที่ดีได้13 เปิดประตูห้องบ่อย ๆ ไม่แนะนำนี้อาจมีผลต่ออุณหภูมิกู้คืน14. หน่วยต้องถูกเก็บไว้จากการระเหย ไวไฟ ระเบิดของเหลวหรือก๊าซ พฤษภาคมนี้สาเหตุความเสียหายอย่างใดอย่างหนึ่งกับหน่วย หรือขาดคุณสมบัติรวมถึงสิ่งอำนวยความสะดวกและอาคาร15. ในกรณีของความผิดปกติหรือกลิ่นเขียน หน่วยต้องได้ทันทีเพิ่มสำหรับการตรวจสอบ16. ถ้าตกเบาบางใกล้ไกล ทันทีปิดไฟ ฟ้าผ่าจากพายุอาจทำให้เกิดไฟไหม้หรือไฟฟ้าช็อต17.แนะนำความปลอดภัยข้างต้นต้องปฏิบัติตามในระหว่างการดำเนินงานของหน่วย ทุกคนปฏิบัติการหน่วยนี้ควรทบทวนคู่มือนี้ ไม่มีค่าตอบแทนหรือเรียกร้องได้กับถ้าเกิดความเสียหายให้กับบุคลากรหรือสิ่งอำนวยความสะดวกที่ได้จากผู้ผลิตการใช้หน่วยนี้ไม่เหมาะสมห้องปฏิบัติการอุปกรณ์ Pty Ltd. 2 ระดับ INCUBATOR เชคเกอร์ 2013 VER1.012. คำอธิบายหน่วยที่ชุด ZHWY-2102/2112 ยางมะตอยบ่มเพาะวิสาหกิจชั้นสองถูกออกแบบมาสำหรับการเติบโตของจุของสิ่งมีชีวิตชีวภาพภายใต้เงื่อนไขของอุณหภูมิ เวลา และอาการกังวลต่อ เหล่านี้ไมโครโปรเซสเซอร์บ่มเพาะวิสาหกิจสโมสรให้มีคุณลักษณะขั้นสูงรวมถึงดิจิตอล setpoints ใช้งานง่ายซิลิโคน ปลุก overtemperatureและข้อเสนอแนะแก้ไขตนเองควบคุมอุณหภูมิ และสั่นเร็วลักษณะการทำงานโดยสรุปl เซนทริกทริมีประสิทธิภาพไดรฟ์ระบบประกันชีวิตของเรียบ และการเคลื่อนไหวงก ๆ อย่างต่อเนื่องไมโครโปรเซสเซอร์ A P.I.D lควบคุมรักษาความแม่นยำอุณหภูมิและการสั่นspeed.l User-friendly sealed keypadallows easy digital settingon running time, speed, andoptimum temperaturecondition.l Large, easy-to-read LCDdisplay presents a menu offunctional readouts.l Automatic shutoff at end ofstudy via electronic timer, adjustable from 1 minute to 500 hours.l Automatic shutoff of shaker when lid is opened which ensures operator safety.l Non-volatile memory retains setpoints during a power failure and restarts shakerautomatically.l Alarms programmed to warn of setpoints deviation and to indicate when timer operation hasexpired.l A high volume air circulation system maintains optimum temperature uniformity, as well asproviding maximum oxygen transfer.l See-through chamber door provides complete visibility and minimize the need to open thechamber frequently.l Mirror-finished stainless steel interior provides excellent durability and an easy-to-cleansurface.Laboratory Equipment Pty Ltd. 2 LEVEL INCUBATOR SHAKER 2013 VER1.013. Control Panel4. Starting the UnitTurn on the power to the unit by pressing the Main Switch located on the right sideof the unit body, the unit enters into the running mode. All keys on the control panelare activated in the running mode. The LED Display shows the preset temperature,speed, timer and finally, measured temperature. Press the Start/Pause key, the unitwill operate on the preset parameters. Second press the Start/Pause key while the unitได้หยุดทำการ กดคีย์ On/Off พลังงานมากกว่า 3 วินาทีหน่วยจะถูกปิด5. อุณหภูมิควบคุมสามารถตั้งอุณหภูมิของหน่วยจากสภาวะ-20 ° C ถึง 60 องศาเซลเซียส ตั้งค่าอุณหภูมิเสร็จตามขั้นตอนต่อไปนี้:1. กดปุ่มยืนยัน/ชุด จาก นั้นกดปุ่มเพิ่มเพื่อป้อนการรหัสผ่าน 32. กดแป้นอุณหภูมิ หน่วยป้อนเข้าสู่โหมดการตั้งค่าอุณหภูมิสามารถตั้งค่าอุณหภูมิผ่านคีย์เพิ่มขึ้น/ลดลง เพื่อยืนยันการตั้งค่าใหม่ กดคีย์ชุด/ยืนยันตั้งค่าใหม่เขียนทับค่าเก่าในหน่วยความจำห้องปฏิบัติการอุปกรณ์ Pty Ltd. 2 ระดับ INCUBATOR เชคเกอร์ 2013 VER1.01เครื่องทำความเย็นกดคีย์ชุด/ยืนยัน แล้วกดปุ่มเพิ่มเพื่อใส่รหัสผ่าน 9กดคีย์เครื่อง หน่วยป้อนเข้าสู่โหมดการตั้งค่าเครื่องทำความเย็นสามารถตั้งค่าพารามิเตอร์แช่แข็งผ่านคีย์เพิ่มขึ้น/ลดลง เพื่อยืนยันใหม่ตั้งค่า กดคีย์การตั้งค่า/การยืนยัน•ถ้าหน่วยที่จะใช้ในช่วงอุณหภูมิแวดล้อม-5 ° C ถึงสภาวะ + 5° C การสามารถกำหนดพารามิเตอร์การแช่แข็งค่า 0.2•ถ้าหน่วยที่จะใช้ในช่วงอุณหภูมิแวดล้อม-20 ° C ถึงแวดล้อม-5 ° C การสามารถกำหนดพารามิเตอร์การแช่แข็งที่ค่า 0.5•ถ้าหน่วยที่จะใช้ในช่วงอุณหภูมิแวดล้อม + 5° C ถึง 60° C การสามารถกำหนดพารามิเตอร์แช่แข็งมูลค่า 2 เครื่องทำความเย็นที่ถูกยกเลิก6. Overtemperature Alarm
In case the gap between measured temperature and setting temperature excels ± 3°C,
the unit sends out audible and visible alarm automatically, and the yellow alarm
indicator flashes. The platform keeps on shaking while the heating or refrigeration
stops. Press the Temperature key will cancel the alarm for 30 minutes. Reset the
temperature value, the unit exits the alarm mode.
7. Speed Control
To set the desired shaking speed, first press the Set/Confirmation key, then press the
Increase key to enter the password 3. Press the Speed key, the unit enters into speed
setting mode. The desired speed value can be set via the Increase/Decrease key. To
confirm the new setting, press the Set/Confirmation key. Setting of a new value
overwrites the old value in the memory.
8. Timer
When timer function is needed, first press the Set/Confirmation key, then press the
Increase key to enter the password 3. Press the Timer key, the unit enters into timer
setting mode, the desired timer value can be set via the Increase/Decrease key. To
confirm the new setting, press the Set/Confirmation key.
By the time the remaining time reaches 0:00, the platform stops shaking, the unit sends out audible
and visible alarm for 2 minutes. Press the Timer key, the alarm will be muted.
Laboratory Equipment Pty Ltd. 2 LEVEL INCUBATOR SHAKER 2013 VER1.01
9. Defrosting
Operation the unit at a low temperature for a long period of time causes the buildup of
frost, which degrades the freezing performance and may disable the temperature from
being controlled at the setpoint. The defrosting function guarantees the unit to operate
smoothly and continuously in such a case. Defrosting function is set via the following
steps:
• First press the Set/Confirmation key, then press the Increase key to enter the
password 9. Press the Refrigeration key, the unit enters into refrigeration mode
(first stage).
• Press the Refrigeration key again, the unit enters into defrosting timer setting
mode (second stage). Defrosting timer value can be set via the
Increase/Decrease key.
• Press the Refrigeration key once more, the unit enters into defrosting time
period setting mode. Defrosting time period can be set via the
Increase/Decrease key.
• To confirm the new setting, press the Set/Confirmation key.
Defrosting Timer Setting refers to the function by which the unit starts defrosting every given time
period. For example, if the defrosting timer value is set at 1, the unit starts defrosting every 1 hour.
If the defrost timer value is set at 2, the unit starts defrosting every 2 hour. However, if the defrost
timer value is set at 0, the unit does not start defrosting at all.
Defrosting Time Period refers to the time span from the moment the unit starts defrosting till it
exits d
การแปล กรุณารอสักครู่..
ผลลัพธ์ (ไทย) 2:[สำเนา]
คัดลอก!
www.labec.com.au
Pty ห้องปฏิบัติการอุปกรณ์ Ltd
อีเมล์: sales@labec.com.au
Ph: 02 9560 2811 •แฟกซ์: 9560 6131 02
ปี 2014
คู่มือการใช้งานอุปกรณ์ห้องปฏิบัติการ Pty Ltd. 2 ระดับ INCUBATOR SHAKER 2013 Ver1.01 ห้องปฏิบัติการอุปกรณ์PTY จำกัดWWW.LABEC.COM.AU สองชั้นINCUBATOR SHAKER ของผู้ใช้ MANUAL เนื้อหา1 คำแนะนำความปลอดภัย2 คำอธิบาย3 แผงควบคุม4 เริ่มต้นหน่วยที่ 5 การควบคุมอุณหภูมิที่ 6 ปลุก Overtemperature 7 ควบคุมความเร็ว8 ฟังก์ชั่นตั้งเวลา9 ละลายน้ำแข็ง10 หน่วยความจำแบบไม่ลบเลือน11 ข้อมูลจำเพาะทางเทคนิค12 ปัญหาการยิง13 บรรจุรายชื่อห้องปฏิบัติการอุปกรณ์ Pty Ltd. 2 ระดับ INCUBATOR SHAKER 2013 Ver1.01 ขอบคุณมากสำหรับการเลือก ZHWY-2102/2112 ซีรีส์สองชั้นบ่มเพาะอิทธิพล เพื่อให้แน่ใจว่าการทำงานที่ราบรื่นของหน่วยนี้เราอยากจะขอให้คุณอ่านคู่มือการใช้งานนี้อย่างละเอียดและมักจะใช้งานเครื่องเช่นนี้นำโดยคู่มือ ยังเก็บไว้ในคู่มือนี้กับหน่วยสำหรับการอ้างอิงในการบำรุงรักษา ในกรณีที่คุณพบปัญหากับหน่วยนี้โปรดอย่าลังเลที่จะติดต่อกับอุปกรณ์ห้องปฏิบัติการ Pty Ltd. เราจะช่วยให้คุณระบุปัญหาและให้บริการแก้ปัญหาสำหรับคุณ. 1 คำแนะนำความปลอดภัย1 หน่วยจะต้องมีสายอย่างถูกต้องเพื่อความปลอดภัยทั้งผู้ประกอบการหรือสิ่งอำนวยความสะดวก มันเป็นเรื่องที่แนะนำให้ใช้สายเฉพาะเพื่อหลีกเลี่ยงการรบกวนจากอุปกรณ์อื่น ๆ . 2 โปรดตรวจสอบแหล่งจ่ายพลังงานไฟฟ้าในท้องถิ่นเป็นไปตามความต้องการไฟฟ้า240V 50 / 60Hz ของหน่วย. 3 อย่าใช้สายไฟหรือสายพิเศษสำหรับหน่วยนี้ไม่โค้งงอหรือบิดโยงอำนาจสายหรือใช้ความตึงเครียดมากเกินไปมันมิฉะนั้นอาจทำให้เกิดความเสียหายกับตัวเครื่องหรือส่งผลให้เกิดไฟฟ้าช็อตหรือไฟไหม้. 4 อย่าสัมผัสหน่วยด้วยมือเปียกหรือใส่เข้าไปในรายการอื่น ๆ เข้าลมเมื่อหน่วยทำงาน มันอาจจะทำให้เกิดไฟไหม้หรือไฟฟ้าช็อต. 5 ห้ามเคลื่อนย้ายหน่วยเมื่อมันมีการดำเนินงาน. 6 ในกรณีที่มีการเคลื่อนย้ายหน่วยเป็นสิ่งที่จำเป็นโปรดตรวจสอบหน่วยเก็บในแนวตั้ง มิฉะนั้นอาจก่อให้เกิดความเสียหายให้กับระบบทำความเย็นภายใน. 7 มันเป็นวิธีที่ดีเพื่อให้แน่ใจว่าแพลตฟอร์มสั่นมาถึงการหยุดนิ่งสมบูรณ์ก่อนที่จะเปิดประตูห้อง. 8 เมื่อเปิดหรือปิดประตูห้องโปรดจัดการด้วยความระมัดระวังอย่าใช้มากเกินไปแรง. 9 ในกรณีที่ฟิวส์จะต้องมีการเปลี่ยนแหล่งจ่ายพลังงานไฟฟ้าจะต้องมีการตัด. 10 หน่วยต้องวางอยู่บนที่เป็นของแข็งพื้นราบใกล้กับเต้าเสียบไฟฟ้าลงดินให้เพียงพอกวาดล้างในทุกด้านของเครื่องอย่างน้อย 20 ซมสำหรับการระบายอากาศที่เหมาะสม หลีกเลี่ยงสถานที่ภายใต้แสงแดดโดยตรงหรือความร้อนที่อยู่ใกล้การผลิตอุปกรณ์เช่นเตาอบ. 11 หลังจากทำความเย็นอย่างต่อเนื่องเป็นเวลา 10 วันก็คือการปฏิบัติที่ดีในการยกเลิกการชื้นหน่วยด้วยการเปลี่ยนฟังก์ชั่นร้อน. 12 การตรวจสอบปกติเพื่อยึดขวดเป็นสิ่งที่จำเป็นเพื่อให้แน่ใจว่าพวกเขาอยู่ในสถานะที่ดี. 13 เปิดบ่อยของประตูห้องไม่แนะนำเช่นนี้อาจมีผลต่ออุณหภูมิการกู้คืน. 14 หน่วยจะต้องเก็บไว้ห่างจากระเหยไวไฟระเบิดของเหลวหรือก๊าซ ซึ่งอาจก่อให้เกิดความเสียหายทั้งการสูญเสียหรือหน่วยให้บริการรวมถึงสิ่งอำนวยความสะดวกและอาคาร. 15 ในกรณีของความผิดปกติหรือกลิ่นเผาไหม้หน่วยจะต้องหยุดทันทีสำหรับการตรวจสอบ. 16 หากวิธีพายุฝนฟ้าคะนองในระยะไกลทันทีปิดไฟ ฟ้าผ่าจากพายุอาจทำให้เกิดไฟไหม้หรือไฟฟ้าช็อต. 17 คำแนะนำด้านความปลอดภัยดังกล่าวข้างต้นจะต้องปฏิบัติตามระหว่างการดำเนินการของหน่วย ทุกคนในการดำเนินงานหน่วยนี้ควรตรวจสอบคู่มือฉบับนี้ ไม่มีการเรียกร้องค่าชดเชยหรือสามารถทำกับผู้ผลิตหากความเสียหายที่เกิดกับบุคลากรหรือสิ่งอำนวยความสะดวกที่เป็นผลมาจากการใช้งานที่ไม่เหมาะสมของหน่วยนี้. อุปกรณ์ห้องปฏิบัติการ Pty Ltd. 2 ระดับ INCUBATOR SHAKER 2013 Ver1.01 2 รายละเอียดหน่วยZHWY-2102/2112 ซีรีส์สองชั้นอิทธิพลบ่มเพาะได้รับการออกแบบสำหรับการเจริญเติบโตความจุขนาดใหญ่ของสิ่งมีชีวิตทางชีวภาพภายใต้เงื่อนไขที่เหมาะสมของอุณหภูมิเวลาและความปั่นป่วน เหล่านี้ไมโครโปรเซสเซอร์ควบคุมอิทธิพลศูนย์บ่มเพาะมีจำนวนคุณสมบัติขั้นสูงรวมถึงsetpoints ดิจิตอลที่ใช้งานง่ายปุ่มกด, ปลุก overtemperature และตนเองแก้ไขข้อเสนอแนะการควบคุมอุณหภูมิและความเร็วในการสั่น. คุณสมบัติสรุปlสามประหลาดที่มีประสิทธิภาพกลไกการไดรฟ์รับประกันอายุการใช้งานที่ราบรื่นและต่อเนื่องสั่นเคลื่อนไหว. lไมโครโปรเซสเซอร์ PID ควบคุมที่แม่นยำรักษาอุณหภูมิและความสั่นความเร็ว. lปุ่มกดปิดผนึกที่ใช้งานง่ายช่วยให้การตั้งค่าดิจิตอลง่ายในการทำงานเวลาความเร็วและอุณหภูมิที่เหมาะสมสภาพ. lขนาดใหญ่และง่ายต่อการอ่านจอแอลซีดีที่มีการจัดแสดงผลเมนูของreadouts ทำงาน. l shutoff อัตโนมัติในตอนท้ายของการศึกษาผ่านทางจับเวลาอิเล็กทรอนิกส์ปรับตั้งแต่ 1 นาทีถึง 500 ชั่วโมง. l shutoff อัตโนมัติของเครื่องปั่นเมื่อฝาเปิดที่มั่นใจความปลอดภัยผู้ประกอบการ. หน่วยความจำไม่ระเหยlยังคง setpoints ระหว่าง ความล้มเหลวของอำนาจและรีสตาร์ทเครื่องปั่นโดยอัตโนมัติ. สัญญาณเตือนภัยlโปรแกรมที่จะเตือนของการเบี่ยงเบน setpoints และเพื่อแสดงให้เห็นการดำเนินการจับเวลาเมื่อได้หมดอายุ. lปริมาณสูงระบบการไหลเวียนของอากาศรักษาความสม่ำเสมอของอุณหภูมิที่เหมาะสมเช่นเดียวกับการให้การถ่ายโอนออกซิเจนสูงสุด. lดู ประตูห้องทะลุให้การมองเห็นที่สมบูรณ์และลดความจำเป็นที่จะต้องเปิดในห้องบ่อย. lภายในสแตนเลสกระจกสำเร็จรูปให้ความทนทานที่ดีเยี่ยมและง่ายต่อการทำความสะอาดพื้นผิว. อุปกรณ์ห้องปฏิบัติการ Pty Ltd. 2 ระดับ INCUBATOR SHAKER 2013 Ver1.01 3 แผงควบคุม4 เริ่มต้นหน่วยเปิดเพาเวอร์ไปยังหน่วยโดยการกดสวิทช์หลักตั้งอยู่บนด้านขวาของร่างกายหน่วยหน่วยเข้าสู่โหมดการทำงาน ทุกปุ่มบนแผงควบคุมจะเปิดใช้งานในโหมดการทำงาน จอแสดงผล LED แสดงอุณหภูมิที่ตั้งไว้ความเร็วจับเวลาและสุดท้ายวัดอุณหภูมิ กดปุ่ม Start / หยุดชั่วคราวที่สำคัญหน่วยงานที่จะดำเนินการกับพารามิเตอร์ที่กำหนดไว้ล่วงหน้า กดที่สองเริ่มต้น / หยุดชั่วคราวที่สำคัญในขณะที่หน่วยงานที่กำลังทำงานอยู่หยุดการดำเนินการ โฮลดิ้งเปิด / ปิดที่สำคัญมานานกว่า 3 วินาทีหน่วยขับเคลื่อนปิด. 5 การควบคุมอุณหภูมิอุณหภูมิของเครื่องสามารถตั้งค่าจากรอบ-20 ° C ถึง 60 ° C การตั้งค่าอุณหภูมิจะแล้วเสร็จในขั้นตอนต่อไปนี้: 1 กดตั้ง / การยืนยันที่สำคัญแล้วกดปุ่มเพิ่มเพื่อป้อนรหัสผ่าน3. 2 กดปุ่มอุณหภูมิหน่วยเข้าสู่โหมดการตั้งค่าอุณหภูมิ. ค่าอุณหภูมิที่สามารถตั้งค่าผ่านทางเพิ่มขึ้น / ลดลงที่สำคัญ เพื่อยืนยันการตั้งค่าใหม่ให้กดตั้ง / การยืนยันที่สำคัญ. การตั้งค่าของค่าใหม่เขียนทับค่าเดิมในหน่วยความจำ. อุปกรณ์ห้องปฏิบัติการ Pty Ltd. 2 ระดับ INCUBATOR SHAKER 2013 Ver1.01 ทำความเย็นกดตั้ง / การยืนยันที่สำคัญจากนั้นกด เพิ่มกุญแจสำคัญในการป้อนรหัสผ่าน 9. กดปุ่มเครื่องทำความเย็นที่หน่วยเข้าสู่โหมดการตั้งค่าเครื่องทำความเย็น. พารามิเตอร์ทำความเย็นที่สามารถตั้งค่าผ่านทางเพิ่มขึ้น / ลดลงที่สำคัญ เพื่อยืนยันใหม่การตั้งค่าให้กดตั้ง / การยืนยันที่สำคัญ. •หากหน่วยที่จะใช้ในช่วงอุณหภูมิโดยรอบ-5 ° C ที่จะล้อมรอบ + 5 ° C, พารามิเตอร์ทำความเย็นที่สามารถตั้งค่าด้วยมูลค่า 0.2. • หากหน่วยงานที่จะนำมาใช้ในช่วงอุณหภูมิโดยรอบ 20 ° C ถึงล้อมรอบ-5 ° C, พารามิเตอร์ทำความเย็นที่สามารถตั้งค่าด้วยมูลค่า 0.5. •หากหน่วยที่จะใช้ในช่วงที่อุณหภูมิโดยรอบ + 5 ° C ถึง 60 ° C, พารามิเตอร์ทำความเย็นที่สามารถตั้งค่าในราคาที่ 2. ทำความเย็นจะถูกยกเลิก. 6 ปลุก Overtemperature ในกรณีที่ช่องว่างระหว่างอุณหภูมิวัดอุณหภูมิและการตั้งค่า excels ± 3 ° C, หน่วยส่งออกปลุกเสียงและมองเห็นได้โดยอัตโนมัติและการเตือนภัยสีเหลืองกะพริบ แพลตฟอร์มช่วยในการสั่นในขณะที่ความร้อนหรือความเย็นหยุด กดปุ่มอุณหภูมิจะยกเลิกการเตือนภัยเป็นเวลา 30 นาที รีเซ็ตค่าอุณหภูมิหน่วยออกจากโหมดการเตือนภัย. 7 การควบคุมความเร็วในการตั้งค่าที่ต้องการสั่นความเร็วกดแรกตั้ง / การยืนยันที่สำคัญจากนั้นกดที่สำคัญเพิ่มขึ้นให้ใส่รหัสผ่าน3. กดปุ่มความเร็วหน่วยเข้าสู่ความเร็วโหมดการตั้งค่า ค่าความเร็วที่ต้องการสามารถตั้งค่าผ่านทางเพิ่มขึ้น / ลดลงที่สำคัญ เพื่อยืนยันการตั้งค่าใหม่ให้กดตั้ง / การยืนยันที่สำคัญ การตั้งค่าของค่าใหม่เขียนทับค่าเดิมในหน่วยความจำ. 8 ตั้งเวลาเมื่อฟังก์ชันจับเวลาเป็นสิ่งจำเป็นกดแรกตั้ง / การยืนยันที่สำคัญจากนั้นกดที่สำคัญเพิ่มขึ้นให้ใส่รหัสผ่าน3. กดปุ่มจับเวลาหน่วยจับเวลาเข้าสู่โหมดการตั้งค่าค่าที่ต้องการจับเวลาสามารถตั้งค่าผ่านทางเพิ่มขึ้น/ ที่สำคัญลดลง เพื่อยืนยันการตั้งค่าใหม่ให้กดตั้ง / การยืนยันที่สำคัญ. ตามเวลาเวลาที่เหลือถึง 00:00 แพลตฟอร์มหยุดสั่นหน่วยส่งออกเสียงปลุกและสามารถมองเห็นได้เป็นเวลา2 นาที กดปุ่มตั้งเวลาปลุกจะถูกปิด. อุปกรณ์ห้องปฏิบัติการ Pty Ltd. 2 ระดับ INCUBATOR SHAKER 2013 Ver1.01 9 ละลายน้ำแข็งการดำเนินงานหน่วยที่อุณหภูมิต่ำเป็นระยะเวลานานของเวลาทำให้เกิดการสะสมของน้ำค้างแข็งซึ่งลดประสิทธิภาพการแช่แข็งและอาจปิดการใช้งานที่อุณหภูมิจากถูกควบคุมที่SetPoint ฟังก์ชั่นละลายน้ำแข็งรับประกันหน่วยในการดำเนินงานได้อย่างราบรื่นและต่อเนื่องในกรณีเช่นนี้ ฟังก์ชั่นที่ละลายน้ำแข็งมีการตั้งค่าผ่านทางดังต่อไปนี้ขั้นตอน•กดครั้งที่ตั้ง/ การยืนยันที่สำคัญแล้วกดปุ่มเพิ่มเพื่อป้อนรหัสผ่าน9. กดปุ่มเครื่องทำความเย็นที่หน่วยเข้าสู่โหมดทำความเย็น(ขั้นแรก). •กดปุ่มเครื่องทำความเย็น อีกหน่วยเข้าสู่การตั้งค่าการจับเวลาละลายน้ำแข็งโหมด(ขั้นตอนที่สอง) ค่าจับเวลาละลายน้ำแข็งที่สามารถตั้งค่าผ่านทางเพิ่มขึ้น / ลดลงที่สำคัญ. •กดปุ่มทำความเย็นอีกครั้งหนึ่งหน่วยเข้าสู่เวลาละลายน้ำแข็งระยะเวลาโหมดการตั้งค่า ช่วงเวลาละลายน้ำแข็งที่สามารถตั้งค่าผ่านทางเพิ่มขึ้น / ลดลงที่สำคัญ. •เพื่อยืนยันการตั้งค่าใหม่ให้กดตั้ง / การยืนยันที่สำคัญ. ละลายน้ำแข็งตั้งค่าตัวตั้งเวลาหมายถึงฟังก์ชั่นโดยที่หน่วยเริ่มละลายน้ำแข็งทุกครั้งที่ได้รับระยะเวลา ตัวอย่างเช่นถ้าค่าจับเวลาละลายน้ำแข็งมีการตั้งค่าที่ 1, หน่วยเริ่มละลายน้ำแข็งทุก 1 ชม. ถ้าค่าจับเวลาละลายน้ำแข็งมีการตั้งค่าที่ 2, หน่วยเริ่มละลายน้ำแข็งทุก 2 ชั่วโมง แต่ถ้าละลายน้ําแข็งค่าจับเวลาตั้งไว้ที่ 0, หน่วยไม่เริ่มละลายน้ำแข็งที่ทุกคน. ละลายน้ำแข็งช่วงเวลาหมายถึงช่วงเวลาจากช่วงเวลาที่หน่วยเริ่มละลายน้ำแข็งจนออก d


















































































































































































การแปล กรุณารอสักครู่..
ผลลัพธ์ (ไทย) 3:[สำเนา]
คัดลอก!
www.labec . com au
อุปกรณ์ห้องปฏิบัติการ Pty Ltd
อีเมล์ : ขาย @ labec . com au
M : 02 9560 2811 โทรสาร : 02 - 9560 6131 2014


คู่มือการใช้งานอุปกรณ์ห้องปฏิบัติการ Pty จำกัดระดับ 2 ตู้ Shaker 2013 อุปกรณ์ห้องปฏิบัติการ ver1.01


www.labec Pty Ltd . com au

สองระดับ คู่มือผู้ใช้ตู้ Shaker

เนื้อหา
1 ความปลอดภัย
2 รายละเอียด
3 ใน ' แผงควบคุม '
4 เริ่มจากหน่วย
5 ควบคุมอุณหภูมิด้วย
6overtemperature ปลุก
7 การควบคุม
ความเร็ว 8 ฟังก์ชันจับเวลา
9 การละลายน้ำแข็ง
10 การหลีกเลี่ยงภาษี
11 ข้อมูลทางเทคนิค
12 ปัญหาในการถ่ายภาพ
13 รายการบรรจุ
อุปกรณ์ห้องปฏิบัติการ Pty จำกัดระดับ 2 ตู้ Shaker 2013 ver1.01
ขอบคุณมากสําหรับการเลือก zhwy-2102 / 2112 ชุด 2 ชั้น ตู้ฟักไข่
สโมสร . เพื่อให้แน่ใจว่าการดำเนินงานที่ราบรื่นของหน่วยนี้ เราอยากจะขอให้คุณอ่าน
การสอนคู่มือฉบับนี้อย่างละเอียด และมักใช้เครื่องเป็นแนวทางโดยคู่มือเล่มนี้

ยังเก็บคู่มือกับหน่วยซ่อมบำรุงสำหรับการอ้างอิง ในกรณีที่คุณพบปัญหากับ
หน่วยนี้ กรุณารู้สึกฟรีเพื่อติดต่อไปยังอุปกรณ์ห้องปฏิบัติการ
เราจะช่วยให้คุณระบุปัญหาและให้บริการถ่ายภาพปัญหา
.
1 ความปลอดภัย
1หน่วยจะต้องถูกโอนให้ความปลอดภัยให้กับผู้ประกอบการหรือสิ่งอำนวยความสะดวก มันเป็นขอแนะนำให้ใช้บรรทัด
ทุ่มเทเพื่อหลีกเลี่ยงสัญญาณรบกวนจากอุปกรณ์อื่น ๆ .
2 โปรดตรวจสอบแหล่งจ่ายไฟไฟฟ้าท้องถิ่นเป็นไปตามความต้องการ
ไฟฟ้า 240V 50 / 60Hz ของหน่วย .
3 อย่าใช้สายไฟหรือสายพิเศษสำหรับหน่วยนี้ ไม่งอ พาดพิง หรือบิดพลัง
สายไฟหรือใช้แรงมากเกินไปนั้น มิฉะนั้นอาจก่อให้เกิดความเสียหายต่อหน่วย ส่งผลให้ไฟฟ้าดูด หรือไฟ
.
4 ไม่สัมผัสหน่วยด้วยมือที่เปียกหรือแทรกรายการอื่น ๆลงในลมเข้า เมื่อเป็นหน่วย
วิ่ง มันอาจทำให้เกิดไฟไหม้หรือไฟฟ้าช็อตได้
5 ไม่ย้ายหน่วยเมื่อมันเป็นปฏิบัติการ .
6 ในกรณีย้ายหน่วยที่จำเป็นโปรดตรวจสอบหน่วยเก็บตรงมิฉะนั้นอาจ
ให้เกิดความเสียหายกับระบบทำความเย็นภายใน .
7 มันคือการปฏิบัติที่ดีเพื่อให้แน่ใจว่าสั่นแพลตฟอร์มมาหยุดนิ่งที่สมบูรณ์ก่อนที่จะเปิดประตูห้อง
.
8 เมื่อเปิดหรือปิดประตูห้อง ช่วยจัดการกับการดูแลที่ไม่ต้องใช้แรงมาก
.
9 ในกรณีที่ต้องมีการเปลี่ยนฟิวส์ , ไฟฟ้าจัดหาได้ถูกตัดออก
10หน่วยจะต้องถูกวางไว้บนพื้นแข็ง แบน ใกล้กักบริเวณเต้าเสียบให้เคลียร์เพียงพอ
ในทุกด้านของหน่วย อย่างน้อย 20 เซนติเมตร เพื่อระบายอากาศที่เหมาะสม หลีกเลี่ยงสถานที่ภายใต้
แสงแดดโดยตรง หรือใกล้ความร้อน ผลิตอุปกรณ์ เช่น เตาอบ
11 หลังจากแช่แข็งต่อเนื่อง 10 วัน เป็นวิธีปฏิบัติที่ดีที่จะ de ชื้นหน่วยโดยการเปิดในฟังก์ชันความร้อน
.
12ตรวจสอบปกติขวดหนีบเป็นสิ่งที่จำเป็นเพื่อให้แน่ใจว่าพวกเขาอยู่ในสถานะที่ดี
13 ประตูห้องถูกเปิดบ่อยไม่แนะนำนี้อาจส่งผลกระทบต่อการกู้คืนอุณหภูมิ
.
14 หน่วยจะต้องถูกเก็บไว้ห่างจากสารระเหยไวไฟ ระเบิดของเหลวหรือก๊าซ นี้อาจทำให้เกิดความเสียหายให้กับหน่วย
ขาดทุนจากคุณสมบัติรวมถึงอาคารและสิ่งอำนวยความสะดวก .
ที่ 15ในกรณีของความผิดปกติหรือกลิ่นเผาไหม้ หน่วยจะต้องได้ทันทีหยุดการตรวจสอบ .
16 ถ้าฝนฟ้าคะนองเป็นแนวไกลทันทีปิดไฟ สายฟ้าจาก
พายุอาจทำให้เกิดไฟไหม้หรือไฟฟ้าช็อตได้
17 คำแนะนำความปลอดภัยข้างต้นจะต้องปฏิบัติตามในการดำเนินงานของหน่วย ทุกคน
หน่วยปฏิบัติการนี้ควรตรวจสอบนี้คู่มือไม่มีการชดเชย หรือเรียกร้องสิทธิสามารถทำได้กับ
ผู้ผลิตหากความเสียหายที่เกิดกับบุคลากรหรือเครื่องที่เป็นผลจากการใช้ที่ไม่เหมาะสมของหน่วยนี้
.
อุปกรณ์ห้องปฏิบัติการ Pty จำกัด 2 ระดับ 2013 ver1.01 shaker incubator
2 หน่วยรายละเอียด
zhwy-2102 / 2112 ชุด 2 ชั้น ตู้สโมสรถูกออกแบบมาสำหรับ
การเจริญเติบโตความจุขนาดใหญ่ของสิ่งมีชีวิตทางชีวภาพภายใต้สภาวะของอุณหภูมิ เวลา และเขย่า เหล่านี้

เอไมโครโปรเซสเซอร์ควบคุมสโมสรเสนอคุณลักษณะขั้นสูงรวมถึงหมายเลข

ใช้ง่ายปุ่มกด setpoints

ปลุกดิจิตอล overtemperature และตนเองแก้ไขความคิดเห็น

สั่นควบคุมอุณหภูมิและความเร็ว ในการสรุป
L
คุณสมบัติที่มีประสิทธิภาพสามกลไกรับประกัน

ขับรถนอกรีตชีวิตที่เรียบและ

ผมอย่างต่อเนื่องสั่นไหว เป็น p.i.d ไมโครโปรเซสเซอร์ควบคุมอุณหภูมิแม่นยำและรักษา



ผมเขย่าความเร็ว ผู้ใช้ที่เป็นมิตรช่วยให้ง่ายต่อการตั้งค่าปุ่มกดดิจิตอล ผนึก

ในงานเวลา , ความเร็ว , อุณหภูมิที่เหมาะสมและ

.
L ขนาดใหญ่อ่านง่ายแสดงผล LCD
นำเสนอเมนู

l ผลการคำนวณที่ตามมาในการทํางาน อัตโนมัติ shutoff ที่จบการศึกษาผ่าน
จับเวลาอิเล็กทรอนิกส์
การแปล กรุณารอสักครู่..
 
ภาษาอื่น ๆ
การสนับสนุนเครื่องมือแปลภาษา: กรีก, กันนาดา, กาลิเชียน, คลิงออน, คอร์สิกา, คาซัค, คาตาลัน, คินยารวันดา, คีร์กิซ, คุชราต, จอร์เจีย, จีน, จีนดั้งเดิม, ชวา, ชิเชวา, ซามัว, ซีบัวโน, ซุนดา, ซูลู, ญี่ปุ่น, ดัตช์, ตรวจหาภาษา, ตุรกี, ทมิฬ, ทาจิก, ทาทาร์, นอร์เวย์, บอสเนีย, บัลแกเรีย, บาสก์, ปัญจาป, ฝรั่งเศส, พาชตู, ฟริเชียน, ฟินแลนด์, ฟิลิปปินส์, ภาษาอินโดนีเซี, มองโกเลีย, มัลทีส, มาซีโดเนีย, มาราฐี, มาลากาซี, มาลายาลัม, มาเลย์, ม้ง, ยิดดิช, ยูเครน, รัสเซีย, ละติน, ลักเซมเบิร์ก, ลัตเวีย, ลาว, ลิทัวเนีย, สวาฮิลี, สวีเดน, สิงหล, สินธี, สเปน, สโลวัก, สโลวีเนีย, อังกฤษ, อัมฮาริก, อาร์เซอร์ไบจัน, อาร์เมเนีย, อาหรับ, อิกโบ, อิตาลี, อุยกูร์, อุสเบกิสถาน, อูรดู, ฮังการี, ฮัวซา, ฮาวาย, ฮินดี, ฮีบรู, เกลิกสกอต, เกาหลี, เขมร, เคิร์ด, เช็ก, เซอร์เบียน, เซโซโท, เดนมาร์ก, เตลูกู, เติร์กเมน, เนปาล, เบงกอล, เบลารุส, เปอร์เซีย, เมารี, เมียนมา (พม่า), เยอรมัน, เวลส์, เวียดนาม, เอสเปอแรนโต, เอสโทเนีย, เฮติครีโอล, แอฟริกา, แอลเบเนีย, โคซา, โครเอเชีย, โชนา, โซมาลี, โปรตุเกส, โปแลนด์, โยรูบา, โรมาเนีย, โอเดีย (โอริยา), ไทย, ไอซ์แลนด์, ไอร์แลนด์, การแปลภาษา.

Copyright ©2025 I Love Translation. All reserved.

E-mail: