A major advantage of the DME is that each drop has a smooth and uncontaminated surface free from any adsorbed analyte or impurity. The self-renewing electrode does not need to be cleaned or polished like a solid electrode. This advantage comes at the cost of a working electrode with a constantly changing surface area. Since the drops are produced predictably the changing surface area can be accounted for or even used advantageously. In addition the drops' growth causes capacitive current and faradaic current. These changing current effects combined with experiments where the potential is continuously changed can result in noisy traces. In some experiments the traces are continually sampled, showing all the current deviation resulting from the drop growth. Other sampling methods smooth the data by sampling the current at the electrode only once per drop at a specific size. The DME's periodic expansion into the solution and hemispherical shape also affects the way the analyte diffuses to the electrode surface. The DME consists of a fine capillary with a bore size of 20-50 µm.