We made guidance for installation of AT-710 on Thai Sanmei on coming 14th. Dec. 2015.
Please note below points.
1. Please bring your note PC.
2. Please save attached "AT-710.zip" on your PC, open and copy & save the files on your USB.
The zip file is 3 methods parameter for this customer.
Method 21: Nickel chloride in plating solution (ETIQ-99321C) with C373
Method 22: Quantitative determination of boric acid(ETIQ-96082) with C171
Method 23: Nickel in Nickel Plating Solution (ETIQ-96082) with photometric sensor
3. We send you application notes they want to do by our AT-710S. They also are doing manual titration now.
Please talk with customer about each consumption volume(mL) of titrant and change the "max volume".
Regarding Method 23, we make "Dose" mode which 10mL dispense of EDTA before finding end-point.
If the end-point (mL) is located before 10mL, please remove the "Dose" mode.
4. If you need manual for AT-710, please download from below link .
This installation is very important since we have another inquiry from Tanabe Thailand which Mr. Segawa introduced.
Please have a imagine for this installation and if you have some questions, please do not hesitate to contact us.