5-9
4. Follow the on-screen instructions. Between each bottle change, rinse the filters by dipping them into distilled water
(approximately 50 ml) to avoid cross-contamination of the solutions.
5. At the end of the wizard, the “What would you like to do?” dialog appears. Click the Shutdown Instrument, Standby,
or Eject Chip button (as applicable).
5.5 Collection Tank
1. Check the collection tube fittings at the instrument and collection tank ends.
2. Verify that the tubing is not pinched or clogged,
3. Verify that the collection tank is placed below the instrument reservoir.
4. Empty the collection tank and properly dispose of the contents.