image noise was 43% of our usual clinical
value. Thus, using the constant noise paradigm, image noise would be unacceptably high for
small patients compared with the clinical requirements reflected by our existing technique
charts. In addition, the requirement of constant noise would overexpose large patients
compared with what we currently use, delivering increased dose to achieve lower noise
values than what we have found to be clinically necessary.