Validation Assays. To verify that the analytical method complies
with criteria for the performance characteristics as established by
Commission Decision 2002/657/EC, the following described
validation procedures were carried out.
Linearity was evaluated through a calibration curve assembled by
plotting each PAH metabolite peak area versus the concentration of
standard solutions, at five concentration levels. The concentration
levels ranged between 750 and 2000 ng/mL for 1-hydroxynaphthalene,
between 30 and 75 ng/mL for 2-hydroxyfluorene, and between 150
and 300 ng/mL for 1-hydroxypyrene. Linearity was also evaluated in
matrix-matched samples at 750, 850, 1000, 1500, and 2000 ng/g; 30,
40, 50, 60, and 75 ng/g; and 150, 175, 200, 250, and 300 ng/g 1-
hydroxynaphthalene,2-hydroxyfluorene, and 1-hydroxypyrene, respectively
Assays ตรวจสอบ เพื่อตรวจสอบว่า วิธีการวิเคราะห์ตาม
เกณฑ์สำหรับลักษณะของประสิทธิภาพการทำงานก่อตั้งขึ้นโดย
กรรมการตัดสิน 2002/657/EC ต่อไปนี้อธิบาย
ขั้นตอนตรวจสอบได้จำหน่ายออก
แบบดอกไม้ถูกประเมิน โดยการเทียบเส้นโค้งประกอบโดย
พล็อตแต่ละพื้นที่สูงสุด metabolite ละเมื่อเทียบกับความเข้มข้นของ
แก้ ที่ระดับความเข้มข้น 5 ความเข้มข้น
ระดับที่อยู่ในช่วงระหว่าง 2000 และ 750 ng/mL สำหรับ 1-hydroxynaphthalene,
ระหว่าง 30 และ 75 ng/mL ใน 2 hydroxyfluorene และ 150
และ ng 300 มล.ใน 1 hydroxypyrene แบบดอกไม้ยังถูกประเมินใน
ตัวอย่างตรงกับเมตริกซ์ ที่ 750, 850, 1000, 1500, 2000 ng/g 30,
40, 50, 60 และ 75 ng/g และ 150, 175, 200, 250 และ 300 ng/g 1-
hydroxynaphthalene2 hydroxyfluorene และ 1-hydroxypyrene ตามลำดับ
การแปล กรุณารอสักครู่..

Validation Assays. To verify that the analytical method complies
with criteria for the performance characteristics as established by
Commission Decision 2002/657/EC, the following described
validation procedures were carried out.
Linearity was evaluated through a calibration curve assembled by
plotting each PAH metabolite peak area versus the concentration of
standard solutions, at five concentration levels. The concentration
levels ranged between 750 and 2000 ng/mL for 1-hydroxynaphthalene,
between 30 and 75 ng/mL for 2-hydroxyfluorene, and between 150
and 300 ng/mL for 1-hydroxypyrene. Linearity was also evaluated in
matrix-matched samples at 750, 850, 1000, 1500, and 2000 ng/g; 30,
40, 50, 60, and 75 ng/g; and 150, 175, 200, 250, and 300 ng/g 1-
hydroxynaphthalene,2-hydroxyfluorene, and 1-hydroxypyrene, respectively
การแปล กรุณารอสักครู่..

การตรวจสอบ ) . เพื่อตรวจสอบว่าวิธีการวิเคราะห์ที่สอดคล้องกับเกณฑ์
ลักษณะการปฏิบัติงานที่จัดตั้งขึ้นโดยคณะกรรมการการตัดสินใจ
2002 / 657 / EC ต่อไปนี้อธิบายขั้นตอนการตรวจสอบ
.
การตรวจวัดประเมินผ่านเป็นรูปโค้งประกอบโดย
พล็อตแต่ละพื้นที่ป่า อุณหภูมิ สูงสุดเมื่อเทียบกับความเข้มข้นของ
โซลูชั่นมาตรฐาน5 ระดับความเข้มข้น ระดับความเข้มข้น
อยู่ระหว่าง 750 และ 2000 ng / ml 1-hydroxynaphthalene
ระหว่าง 30 และ 75 , ng / ml 2-hydroxyfluorene ระหว่าง 150 และ 300 ng / ml
1-hydroxypyrene . การตรวจวัดและประเมินในตัวอย่างที่ตรงกับ
Matrix 750 , 850 , 1000 , 1500 และ 2000 ng / g ;
30 , 40 , 50 , 60 , 75 กรัม / กรัม และ 150 , 175 , 200 , 250 และ 300 ng / g -
hydroxynaphthalene 1 ,2-hydroxyfluorene และ 1-hydroxypyrene ตามลำดับ
การแปล กรุณารอสักครู่..