This work reports the fabrication of disposable three-electrode cells with integrated sputtered metal-film electrodes. Theworking electrodewas a bismuth-film electrode (BiFE) while the reference and counter electrodes were made of Ag and Pt, respectively. The deposition of the metal layers was carried out by sputtering of the respective metals on a silicon substrate while the exact geometry of the electrodes was defined via a metal mask placed on the substrate during the deposition process. Initially, the electrodes were characterised by cyclic voltammetry. The utility of these devices was tested for the trace determination of Ni(II) by square wave adsorptive stripping voltammetry (SWAdSV) after complexation with dimethylglyoxime (DMG). The experimental variables (the presence of oxygen, the DMG concentration, the preconcentration potential, the accumulation time and the SW parameters), aswell as potential interferences,were investigated. Using the selected conditions, the 3 limit of detection was 100 ng L−1 for Ni(II) (for 90 s of preconcentration) and the relative standard deviation for Ni(II) was 2.3% at the 10gL−1 level (n = 8). Finally, the method was applied to the determination of Ni(II) in a certified river water sample.