1. Place the roller clamp of the IV set at the bottom of the pump and close the roller clamp.
2. Put the injecting needle into the IV solution container and fill the solution in the chamber to between 1/3 and 1/2.
3. Open the roller clamp to remove the air from the tube and then close the roller clamp.
4. Mount the IV solution container on the pole. The solution container must not be placed more than 1 m above the patient’s heart.
5. Push the door knob upwards to open the door.
6. Press the ⑥ auxiliary door lock to open the ③ pumping unit.
7. Hold ① with a hand and gently stretch in the direction of
⑤.
8. Press the IV set tightly onto the ② air detector to keep the tube in position.
9. Hold ⑤ with a hand so as not to stretch ③ and press the
④ Occlusion Detector to hold the tube in position.
10. The Air Alarm or the Occlusion Alarm may be triggered if the IV set is not properly installed.
11. Press the door knob downwards to close the door. The auxiliary door will automatically close with door.
12. Open the roller clamp in the IV set.
13. Remove the air in the tube and check if there is leakage prior to putting the needle into the patient’s vein.