Calibration
5.1 Standard curves must be established (as outlined in Procedure 8.3) for each
photometer used, and are not interchangeable. For color values less than 250, a 5 cm cell path is recommended. Less than 5 cm cell paths may be used if calibration is performed with the shorter cell.