10.4.2 Frequency. Prior to the analyses of any samples, blanks, or calibration standards, the Laboratory
must establish that the GC/MS system meets the mass spectral ion abundance criteria for the instrument
performance check standard containing BFB. The instrument performance check solution must be analyzed
initially and once per 24-hour time period of operation.