• Press the Table button on the Program toolbar. This will bring up the Liquid Analysis window.
• In row 1 of the table, double click on the Description field, which will bring up a dialog box with two choices, BLANK and Std__. Click on BLANK to choose this description. Be careful to follow this step carefully. The BLANK and Std__ descriptions need to be inserted exactly into the Description field or the computer software will be unable to correctly recognize them.
• In the V, ml field type 1.
• Click Graph on the Table toolbar. This will return you to the integration window.
• Repeat the baseline check step: Click the Baseline Check button on the Control toolbar. Wait 15 seconds and then click the button again to establish baseline.
• Press Start in the Integration window then inject 1 ml of the 1 g/L standard into impinger # 1. Wait 60 seconds and press End.