The McLeod gauge is independent of gas composition. If, however, the gas contains condensable material and during compression it condenses, the reading of the gauge is faulty. The gauge is not capable of continuous reading and the scale is of square law type. For linearizing the scale at comparatively higher pressures, a second volume is introduced as shown in the figure below, where the scale shown is linear.