Selected samples were analyzed by full scan GC–MS to confirm
the GC-lECD results. An Agilent System 6890 series Plus gas chromatograph
(USA), equipped with an Agilent 5973 mass selective
detector, an Agilent 7683 series Autosampler and a split/splitless
capillary injector port, was employed. The column used was
HP-5 MS (Aglient, Folsom, CA) capillary column of 30 m, 0.25 mm
id., 0.25 lm film thickness. The carrier gas was helium at a flow
rate of 0.5 ml min1. Inlet temperature was 225 C with injection
volume of 1 lL. The following oven temperature programme was
adopted: 60 C (2 min), 10 C min1 to 160 C and finally 3 C
min1 to 260 C (20 min). The carrier gas (He) flow rate was kept
in constant flow mode at 1.0 mL min1. Sample injection was
carried out splitless at 240 C with 1 min purge off. GC–MS interface
was 280 C. Chemistation software was used for instrument
control and data analysis
ตัวอย่างวิเคราะห์โดย GC และสแกนเต็มรูปแบบ MS ยืนยัน
ผล lecd GC . เป็นระบบด้าน 6890 ชุดบวกแก๊ส
( USA ) , อุปกรณ์ที่มีเลนต์ 5973 มวล selective
เครื่องตรวจจับ , เลนต์ 7683 ชุด autosampler และแยก / splitless
ฝอยฉีดพอร์ตที่ใช้ คอลัมน์ที่ใช้อยู่
hp-5 MS ( aglient ฟอลซัม , แคลิฟอร์เนีย , คอลัมน์ capillary ) 30 เมตร , 0.25 มม.
ID , 0 .ความหนาของฟิล์ม 25 ไมโครเมตร บริษัทขนส่งแก๊สฮีเลียมที่ไหล
อัตรา 0.5 มิลลิลิตรต่อนาที 1 อุณหภูมิขาเข้าคือ 225 C ฉีด
เล่ม 1 จะ รายการต่อไปนี้คืออุณหภูมิเตาอบ
บุญธรรม : 60 C ( 2 นาที ) , 10 C มิน 1 160 C และสุดท้าย 3 C
มิน 1 260 C ( 20 นาที ) แก๊สตัวพา ( เขา ) อัตราการไหลที่ถูกเก็บไว้
ในโหมดไหลคงที่ที่ 1.0 มิลลิลิตรต่อนาที 1 การฉีดตัวอย่างคือ
ดำเนินการ splitless ที่ 240 องศาเซลเซียส 1 นาทีล้างออก GC – MS อินเตอร์เฟซ
คือ 280 C chemistation ซอฟต์แวร์ที่ใช้ควบคุมอุปกรณ์
และการวิเคราะห์ข้อมูล
การแปล กรุณารอสักครู่..