PREFACEPREFACE
Thank you for selecting the pH 1100/ 2100 benchtop meter. This meter measure pH, millivolts, relative millivolts and temperature. The pH 2100 meter also measures ion concentration. The instruction manual serves to explain the use of the pH 1100/ 2100 bench meter as a step-by-step operational guide to help you familiarize with the meters features and functions. It is structured sequentially with illustration of diagrams that explains the various functions and setup menus available. This manual is written to cover as many anticipated applications and uses of the pH 1100/ 2100 Bench meter as possible. If there are doubts in the use of the meter, please do not hesitate to contact the nearest Authorized Distributor or Customer Service Dept. for assistance. Eutech Instruments/ Oakton Instruments reserve the rights to change, make improvement and modify specifications without prior notice and cannot accept any responsibility for damage or malfunction to the instrument caused by improper use.
Copyright 2002 Eutech Instruments Pte. Ltd. Oakton Instruments All rights reserved. Revision 2.2, Nov 02.
TABLE OF CONTENTS
1 INTRODUCTION 1.1 1.2 1.3 1.Introducing the Bench Meter Series Keypad Rear Panel Electrode Holder 35
STARTING UP 2.1 Back Panel Connections 2.2 Powering up and powering down 2.2.1 Powering up 2.2.2 Powering down
CALIBRATION & MEASUREMENT
3.1 pH calibration 3.1.1 Automatic Temperature Compensation (ATC) 3.1.2 Starting pH calibration 3.1.3 Standard pH buffer calibration 3.1.4 Custom pH buffer calibration 3.1.5 Calibration error messages 3.2 pH measurement 3.2.1 Automatic temperature compensation 3.2.2 Adjusting manual temperature compensation 3.2.3 Taking pH Measurements 4 mV CALIBRATION AND MEASUREMENT 4.1 mV Calibration 4.1.1 mV Calibration error message 4.2 mV Measurement 5 ION CALIBRATION AND MEASUREMENT (ONLY APPLICABLE FOR pH 2100) 5.1 5.2 5.Ion calibration Ion calibration error messages Ion measurement
MEMORY FUNCTIONS 6.1 Memory input 6.1.1 Manually storing a reading into memory 6.2 Memory recall 6.2.1 Recalling readings in manual recall mode 6.2.2 Recalling readings in automatic recall mode
STABILITY INDICATOR ALARM FUNCTIONS 8.1 8.2 High and Low measurement alarm Calibration due alarm Temperature Calibration Temperature Calibration Error Messages
TEMPERATURE CALIBRATION 9.1 9.2
10 SETUP MODE 10.1 Setup mode overview 10.1.1 pH/Temperature setup submenus 10.1.2 mV setup submenus 10.1.3 Ion setup submenus (pH 2100 only) 10.1.4 Meter general configuration setup submenus 10.2 pH/ temperature setup mode 10.2.1 Entering pH/temperature setup mode 10.2.2 pH buffer setup program P1.0 10.2.3 pH resolution setup program P1.1 10.2.4 Temperature unit setup program P1.2 10.2.5 pH Measurement Alarm Setup Program P1.3 10.2.6 pH Calibration Due Alarm Setup program P1.4 10.2.7 Stability Setup Program P1.5 10.2.8 View pH calibration data setup program P1.6
10.2.9 pH calibration data reset setup program P1.7 10.3 mV setup mode 10.3.1 Entering mV setup mode 10.3.2 mV measurement alarm setup program P2.0 10.3.3 mV calibration due alarm setup program P2.1 10.3.4 Stability setup program P2.2 10.3.5 View mV calibration data setup program P2.3 10.3.6 mV calibration data reset setup program P2.4 10.4 Ion setup mode (pH 2100 meter only) 10.4.1 Entering ion setup mode 10.4.2 Ion unit setup mode P3.0 10.4.3 Ion measurement alarm setup program P3.1 10.4.4 Ion calibration due alarm setup program P3.2 10.4.5 Stability setup program P3.3 10.4.6 Ion mode setup program P3.4 10.4.7 View ion calibration data setup program P3.5 10.4.8 Ion calibration data reset setup program P3.6 10.5 Meter general configuration setup mode 10.5.1 Entering meter general setup mode 10.5.2 Date/time setup mode P4.0 10.5.3 Backlight setup mode P4.1 10.5.4 Data log setup mode P4.2 10.5.5 Memory recall setup mode P4.3 10.5.6 Communication setup mode P4.4 10.5.7 Data transfer setup mode P4.5 10.5.8 Memory clear setup mode P4.6 10.5.9 Meter reset setup mode P4.ELECTRODE CARE 11.1 11.2 11.3 Electrode Activation Electrode Maintenance Storing pH/ORP electrodes
Connector
ATC INPUT
Function
For phono jack connection from the temperature probe for Automatic Temperature Compensation. The probe should be a 30K thermistor probe. For connecting sensors with a BNC connector to the meter. The meter accepts any pH, ORP or ISE electrode with a BNC connector. Always make sure that the connector is clean and dry. For connection to the pin tip type reference electrode normally used with half cell (mono) type pH or ISE electrodes. For connection to the strip chart recorder. For connection to the ground earth jack (standard pin tip connectors). For connection to the RS232C serial port. For connection of the AC adapter power supply (included).
REF REC GND RS232 DC
Electrode Holder
The electrode holder is included in the same box as the meter.
To attach the electrode holder to the meter: 1. The electrode holder base attached to the bottom of the meter comes in the shipping position. See Figure 3. Use a Phillips screwdriver to remove the screw holding the electrode holder in shipping position. See Figure 3. Slide the electrode base away from the meter until the second screw slot lines up with the original screw hole. Use the screw removed earlier to secure electrode holder base into position. See Figure 3. The electrode holder arm is reversible. If desired, remove screw holding electrode holder base and slide base out of brackets. Slide base into brackets on opposite side and replace screw. To install the electrode arm, turn meter back to the upright position. See Figure 4. Line up the hole on the base of the electrode holder arm with the peg on the electrode holder base. Slide the hole securely onto the electrode holder base. See Figure 4. The electrode arm is now ready to swing into desired position.
Figure 3: Bottom of Meter
Figure 4: Fix electrode holder arm to holder base
STARTING UP
Attention: Do not get water on the BNC connector during operation. Avoid touching the connector with soiled or wet hands.
2.1 Back Panel Connections
Refer back to Figure 2. Make the necessary connections as mentioned in Section 1.3 Rear Panel, pg 8.
2.2.1 1.
Powering up and powering down
3.1.3 1.
Standard pH buffer calibration
When entering the standard calibration mode, the first buffer calibration point for standard pH buffers is 7.000 or 6.860 (depending upon which pH buffer set is selected). The upper display shows the current reading, the lower display will show Auto Buf Scan: 7.000 (or 6.860). The upper display will flash pH until the reading has stabilized. See Figure 16. If the selected buffer value is not within 1.0 of the measured pH value, the Auto pH Scan will continue to change until the pH buffer is within 1.0. When pH reading has stabilized, press ENTER. The lower display flashes *pH Offset Cal Done. See Figure 17. This calibration point is now stored in the meter. If you are performing a one-point calibration, press EXIT. The screen flashes Sys Updating.Please wait.The meter then returns to the pH measurement mode. You may now start taking pH readings. If you are performing a multi-point calibration, go to step 3.
7.001pH 25.4C ATC Auto Buf Scan: 7.000
Figure 16
7.001pH 25.4C ATC *pH Offset Cal Done
Figure 17
ATC 10.010pH 25.4C 2 Point Cal Done *
Figure 18
Rinse the probe with deionized water or a rinse solution, and place it in the next pH buffer. The lower display automatically scans to the next pH buffer solution. Wait for the measured pH value to stabilize. The upper display will flash pH until the reading has stabilized. When pH reading has stabilized, press ENTER. The lower display flashes 2 Point Cal Done * (3 if you have calibrated 3 points,4 if 4 points, etc.). See Figure 18. This calibration point is now stored in the meter. Repeat steps 3-5 for each point you wish to calibrate. When you are finished calibrating, press EXIT. The screen flashes Sys. Updating. Please wait. The meter then returns to the pH measurement mode. You may now start taking pH readings. If you perform a complete 5-point pH calibration, when you calibrate the fifth point and press ENTER, the meter automatically returns to measurement mode. You may now start taking pH readings.
Error: Cal Aborted
Figure 45
The meter is now in Rel mV measurement mode. See Figure 46. If you have not performed a mV calibration or have reset the meter, the display will read Abs. mV. See Figure 47. 2. Dip the electrode into the sample. The sensor or glass bulb of the electrode must be completely immersed into the sample. Stir the electrode gently to create a homogeneous sample. Allow time for the reading to stabilize. The upper display will show the mV reading, and the lower display will show the current month, day and time. See Figure 46.
Figure 46
Figure 47
ION CALIBRATION AND MEASUREMENT (ONLY APPLICABLE FOR pH 2100)
The pH 2100 meter reads ion concentration in ppm, ppt, mg/l, g/l, mmol/l and mol/l units. See Ion setup Program P3.0 (page 43) to select ion concentration units.
Ion calibration
This instrument is capable of up to 5-point custom ion calibration with a minimum of 2-point calibration. Custom calibration values can be any value between 0.001 and 19998. Once the first calibration value is keyed in, all other calibration will automatically be at least one decade apart. For example if your first calibration point is at 0.001, all other calibration points will be at least one decade apart from 0.001 (0.001, 0.01, 0.1,1.0, 10.0, 100, 1000, 10000). For best accuracy calibrate your meter to points with similar concentrations to the solutions you want to test. 1. 2. Connect an Ion Selective Electrode (ISE) to the BNC input connector on the back of the meter. Turn the meter ON. If the meter is not in the Ion measurement mode, press EXIT and use >/CAL to select Ion. The mode selected will flash. See Figure 48. Press ENTER to confirm selection. The meter is now in Ion measurement m
PREFACEPREFACEขอบคุณสำหรับการเลือก pH 1100 / 2100 benchtop เมตร นี้มิเตอร์วัดค่า pH, millivolts, millivolts สัมพัทธ์ และอุณหภูมิ PH 2100 เมตรนอกจากนี้วัดความเข้มข้นของไอออน บริการคู่มืออธิบายการใช้ pH 1100 / 2100 เมตรผู้พิพากษาเป็นแนวปฏิบัติแต่ละขั้นตอนเพื่อช่วยให้คุณคุ้นกับเมตรคุณลักษณะและฟังก์ชันการ มันเป็นโครงสร้างตามลำดับพร้อมภาพประกอบของไดอะแกรมที่อธิบายการตั้งค่าเมนูและฟังก์ชันต่าง ๆ คู่มือนี้เขียนขึ้นเพื่อให้ครอบคลุมเป็นจำนวนมากคาดว่าโปรแกรมประยุกต์ และใช้ของ pH 1100 / 2100 เมตรผู้พิพากษาเป็นไป หากมีข้อสงสัยในการใช้มิเตอร์ โปรดอย่าลังเลที่จะติดต่อฝ่ายบริการลูกค้าหรือผู้จัดจำหน่ายได้รับอนุญาตที่ใกล้ที่สุดสำหรับความช่วยเหลือ Eutech เครื่องมือ / เครื่องมือ Oakton จองสิทธิ์ในการเปลี่ยนแปลง ทำการปรับปรุง และปรับเปลี่ยนรายละเอียดให้ทราบล่วงหน้า และไม่รับผิดชอบใด ๆ สำหรับความเสียหาย หรือทำงานผิดพลาดกับตราสารที่เกิดจากการใช้ไม่เหมาะสมสงวนลิขสิทธิ์ 2002 Eutech เครื่องมือจำกัด จำกัด Oakton เครื่องมือทั้งหมดสงวนลิขสิทธิ์ ปรับปรุง 2.2, 02 novสารบัญแนะนำ 1 1.1 1.2 1.3 1.แนะนำม้าเมตรชุดแป้นพิมพ์ด้านหลังแผงอิเล็กโทรดถือ 35เริ่มต้นอัพ 2.1 กลับแผงเชื่อมต่อ 2.2 เปิดขึ้นและเปิดลง 2.2.1 เปิดขึ้น 2.2.2 เปิดลงปรับเทียบและประเมิน3.1 ค่า pH เทียบ 3.1.1 อัตโนมัติอุณหภูมิค่าตอบแทน (ATC) 3.1.2 เริ่มต้นค่า pH เทียบเป็น 3.1.3 ค่า pH มาตรฐานบัฟเฟอร์เทียบ 3.1.4 เองค่า pH บัฟเฟอร์เทียบ 3.1.5 เทียบข้อผิดพลาดข้อ 3.2 ค่า pH วัด 3.2.1 อุณหภูมิอัตโนมัติแทน 3.2.2 Adjusting อุณหภูมิด้วยตนเองแทน 3.2.3 การค่า pH 4 วัด mV ปรับเทียบและประเมิน 4.1 mV เทียบ 4.1.1 mV เทียบข้อผิดพลาดข้อ 4.2 mV วัด 5 ไอออนเทียบและประเมิน (ใช้สำหรับค่า pH 2100) 5. 5.1 5.2 ไอออนเทียบไอออนเทียบข้อผิดพลาดข้อความวัดไอออนหน่วยความจำหน่วยความจำฟังก์ชัน 6.1 ใส่ 6.1.1 เก็บอ่านเป็นหน่วยความจำ 6.2 หน่วยความจำเรียกคืน 6.2.1 Recalling อ่านในอ่าน Recalling โหมด 6.2.2 เรียกคืนด้วยตนเองในโหมดอัตโนมัติเรียกคืนด้วยตนเองฟังก์ชันแจ้งเตือนบ่งชี้เสถียรภาพ 8.1 8.2 สูงต่ำวัดปลุกเทียบครบกำหนดปลุกอุณหภูมิเทียบอุณหภูมิเทียบข้อผิดพลาดเทียบ 9.1 9.2การตั้งค่าการตั้งค่าการตั้งค่าโหมด 10.1 10 โหมดภาพรวม 10.1.1 ตั้งค่าอุณหภูมิค่า pH/เมนู ย่อย 10.1.2 mV ไอออนตั้งค่าเมนูย่อยในเมนูย่อย 10.1.3 (pH 2100 เท่า) 10.1.4 เมตรทั่วไปกำหนดค่าการตั้งค่าเมนูย่อย 10.2 ค่า pH / pH/อุณหภูมิ โปรแกรมติดตั้งโหมด 10.2.2 pH บัฟเฟอร์การตั้งค่า P1.0 10.2.3 การป้อนโหมดตั้งค่าอุณหภูมิ 10.2.1 โปรแกรมตั้งค่าความละเอียดค่า pH P1.1 10.2.4 อุณหภูมิหน่วยลง P1.2 10.2.5 ดูความมั่นคงการตั้งค่าโปรแกรม P1.5 10.2.8 ค่า pH วัดปลุกตั้งค่าโปรแกรม P1.3 10.2.6 pH เทียบครบกำหนดปลุกลง P1.4 10.2.7 โปรแกรมติดตั้งข้อมูลเทียบ pH P1.610.2.9 ข้อมูลการปรับเทียบค่า pH ใหม่โปรแกรมติดตั้งโปรแกรม P1.7 10.3 mV ตั้งค่าโหมด 10.3.1 Entering mV ตั้งค่าโหมด 10.3.2 mV วัดปลุกตั้ง P2.0 10.3.3 เทียบ mV ครบกำหนดโปรแกรมติดตั้งปลุก P2.1 10.3.4 เสถียรภาพลง P2.2 10.3.5 ดู mV เทียบข้อมูลการตั้งค่าโปรแกรม P2.3 10.3.6 mV เทียบข้อมูลรีเซ็ตการตั้งค่าโปรแกรมที่ P2.4 10.4 ไอออนโหมดการติดตั้ง (ค่า pH 2100 เมตรเท่านั้น) 10.4.1 Entering ไอออนโหมด 10.4.2 ไอออนหน่วยตั้งค่าโหมดการติดตั้ง P3.0 10.4.3 ไอออนวัดปลุกลง P3.1 10.4.4 เทียบไอออนเนื่องปลุกลง P3.2 10.4.5 เสถียรภาพลง P3.3 10.4.6 โปรแกรมติดตั้งโหมดไอออน P3.4 10.4.7 ดูไอออนเทียบข้อมูลการตั้งค่าโปรแกรม P3.5 10.4.8 ไอออนเทียบข้อมูลรีเซ็ตการตั้งค่าโปรแกรม P3.6 10.5 เมตรกำหนดค่าทั่วไปโหมด 10.5.1 Entering เมตรตั้งค่าทั่วไปโหมดการติดตั้ง 10.5.2 โหมดตั้งค่าวัน / เวลา P4.0 10.5.3 โหมดการติดตั้งแสงไฟ P4.1 10.5.4 ข้อมูลเข้าสู่โหมดการติดตั้ง P4.2 10.5.5 โหมดตั้งค่าเรียกคืนหน่วยความจำ P4.3 10.5.6 โหมดการติดตั้งสื่อสาร P4.4 10.5.7 การถ่ายโอนข้อมูลในโหมดการติดตั้ง P4.5 10.5.8 หน่วยความจำล้างการตั้งค่าโหมด P4.6 10.5.9 เมตรรีเซ็ตการตั้งค่าโหมด P4ดูแลไฟฟ้า 11.1 11.2 11.3 เก็บบำรุงรักษาไฟฟ้าเรียกใช้อิเล็กโทรด pH/ORP หุงตตัวเชื่อมต่อป้อน ATCฟังก์ชันสำหรับ phono แจ็คเชื่อมต่อจากโพรบอุณหภูมิสำหรับการชดเชยอุณหภูมิอัตโนมัติ โพรบควรจะเป็นโพรบ thermistor 30K สำหรับการเชื่อมต่อเซนเซอร์กับ BNC เชื่อมต่อกับมิเตอร์ วัดการยอมรับใด ๆ ค่า pH, ORP หรืออิเสะไฟฟ้าพร้อมขั้วต่อแบบ BNC เสมอให้แน่ใจว่า ตัวเชื่อมต่อสะอาด และแห้ง เชื่อมต่อกับปลายขาอิเล็กโทรดอ้างอิงชนิดปกติใช้กับครึ่งเซลล์ชนิด (ขาวดำ) ค่า pH หรืออิเสะหุงต สำหรับการเชื่อมต่อการบันทึกแผนภูมิแถบ สำหรับเชื่อมต่อกับแจ็คโลกล่าง (ขามาตรฐานแนะนำเชื่อมต่อ) สำหรับเชื่อมต่อกับพอร์ตอนุกรม RS232C สำหรับการเชื่อมต่อของ AC อะแดปเตอร์ไฟ (รวม)อ้างอิง REC GND RS232 DCวางอิเล็กโทรดวางอิเล็กโทรดจะรวมอยู่ในกล่องเดียวกันเป็นตัววัดการวางอิเล็กโทรดกับตัววัด: 1. อิเล็กโทรดถือฐานกับด้านล่างของตัววัดมาในตำแหน่งที่จัดส่ง ดูรูปที่ 3 ใช้ไขควง phillips สำหรับเอาสกรูที่ยึดอิเล็กโทรดในส่งตำแหน่ง ดูรูปที่ 3 เลื่อนฐานไฟฟ้าจากมิเตอร์จนช่องสกรูสองบรรทัดกับรูน็อตเดิม ใช้สกรูที่ออกก่อนหน้านี้ปลอดภัยฐานวางอิเล็กโทรดลงในตำแหน่ง ดูรูปที่ 3 ยึดแขนอิเล็กโทรดเป็นตัวอย่าง ถ้าต้องการ เอาสกรูไฟฟ้ายึดฐานและฐานภาพนิ่งนอกวงเล็บ เลื่อนฐานลงในวงเล็บด้านตรงข้าม และแทนสกรู การติดตั้งแขนอิเล็กโทรด เปิดเมตรไปตรงตำแหน่ง ดูรูปที่ 4 บรรทัดค่าหลุมบนฐานของแขนวางอิเล็กโทรดกับตรึงบนยึดไฟฟ้าพื้นฐาน สไลด์หลุมอย่างปลอดภัยบนยึดไฟฟ้าพื้นฐาน ดูรูปที่ 4 ตอนนี้พร้อมที่จะสวิงในตำแหน่งที่ต้องแขนอิเล็กโทรดได้รูปที่ 3: ด้านล่างของวัดรูปที่ 4: แก้ไขไฟฟ้ายึดแขนเพื่อยึดฐานเริ่มต้นขึ้นความสนใจ: ไม่รับน้ำบนขั้วต่อ BNC ในระหว่างดำเนินการ หลีกเลี่ยงการสัมผัสเชื่อมต่อ ด้วยมือ soiled หรือเปียก2.1 แผงหลังเชื่อมอ้างอิงกลับไปถึงรูปที่ 2 ทำให้การเชื่อมต่อที่จำเป็นดังกล่าวในหัวข้อ 1.3 ด้านหลังแผง pg 82.2.1 1เปิดขึ้นและเปิดลงเป็น 3.1.3 1เทียบบัฟเฟอร์มาตรฐาน pHเมื่อเข้าสู่โหมดการปรับเทียบมาตรฐาน ปรับเทียบจุดแรกของบัฟเฟอร์สำหรับบัฟเฟอร์มาตรฐาน pH เป็น 7.000 หรือ 6.860 (ขึ้นอยู่กับที่ตั้งบัฟเฟอร์ pH ไว้) อ่านปัจจุบันแสดงบนจอแสดงผล จอแสดงผลด้านล่างจะแสดงอัตโนมัติ Buf สแกน: 7.000 (หรือ 6.860) แสดงด้านบนจะแฟลชค่า pH จนอ่านได้เสถียร ดูรูปที่ 16 ถ้าค่าบัฟเฟอร์ที่เลือกไม่ได้อยู่ภายใน 1.0 ของค่า pH วัด pH อัตโนมัติสแกนจะทำการเปลี่ยนแปลงจนบัฟเฟอร์ pH ภายใน 1.0 เมื่ออ่านค่า pH ได้เสถียร กด ENTER ด้านล่างแสดงแฟลช * ค่า pH ตรงข้าม Cal ทำ ดูรูปที่ 17 ปรับเทียบจุดนี้ตอนนี้อยู่ในตัววัด ถ้าคุณกำลังทำการปรับเทียบจุดหนึ่ง กดออก หน้าจอกระพริบรอ Sys Updating.Pleaseตัววัดแล้วส่งกลับไปยังโหมดการวัดค่า pH ตอนนี้คุณอาจเริ่มการอ่านค่า pH ถ้าคุณกำลังทำการปรับเทียบจุดหลาย ไปที่ขั้นตอนที่ 37.001pH สแกน Buf อัตโนมัติ ATC C 25.4:7.000รูปที่ 167.001pH 25.4 C ATC * ค่า pH ตรงข้าม Cal ทำรูปที่ 17ATC 10.010pH 25.4 C 2 ชี้ Cal ทำ *รูปที่ 18ล้างโพรบ ด้วยน้ำ deionized หรือโซลูชันล้าง และวางไว้ในบัฟเฟอร์ pH ถัดไป การแสดงผลต่ำกว่าสแกนโซลูชันบัฟเฟอร์ pH ถัดไปโดยอัตโนมัติ รอค่า pH วัดเพื่อรักษาเสถียรภาพ แสดงด้านบนจะแฟลชค่า pH จนอ่านได้เสถียร เมื่ออ่านค่า pH ได้เสถียร กด ENTER แสดงด้านล่างปรากฏขึ้น 2 จุด Cal ทำ * (3 ถ้าคุณได้รับการปรับเทียบจุด 3, 4 ถ้า 4 จุด ฯลฯ .) ดูรูปที่ 18 ปรับเทียบจุดนี้ตอนนี้อยู่ในตัววัด ทำซ้ำขั้นตอนที่ 3-5 แต่ละจุดที่คุณต้องการ เมื่อคุณทำเสร็จแล้วปรับเทียบ กดออก หน้าจอกระพริบ Sys ปรับปรุง กรุณารอสักครู่ ตัววัดแล้วส่งกลับไปยังโหมดการวัดค่า pH ตอนนี้คุณอาจเริ่มการอ่านค่า pH ถ้าคุณทำการเทียบค่า pH 5 จุดสมบูรณ์ เมื่อคุณปรับเทียบจุดห้า และกด ENTER มิเตอร์อัตโนมัติกลับไปยังโหมดวัด ตอนนี้คุณอาจเริ่มการอ่านค่า pHข้อผิดพลาด: Cal ยกเลิก45 รูปตัววัดอยู่ขณะนี้ในโหมดวัด mV Rel ดูรูป 46 ถ้าคุณได้ทำการปรับเทียบ mV หรือจะตั้งค่ามิเตอร์ จอแสดงผลจะอ่านหยุดตาม mV ดูรูป 47 2. จุ่มอิเล็กโทรดที่เป็นตัวอย่าง หลอดแก้วหรือเซ็นเซอร์ของอิเล็กโทรดต้องมีไปเป็นตัวอย่างทั้งหมด ผัดอิเล็กโทรดเบา ๆ เพื่อสร้างตัวอย่างเป็นเนื้อเดียวกัน ทำให้เวลาอ่านหนังสืออยู่ดี แสดงด้านบนจะแสดง mV อ่าน และการแสดงผลด้านล่างจะแสดงเดือนในปัจจุบัน วัน และเวลา ดูรูป 46รูป 4647 รูปไอออนเทียบและประเมิน (ใช้สำหรับค่า pH 2100)PH 2100 เมตรอ่านความเข้มข้นของไอออนในหน่วย ppm, ppt, mg/l, g/l, mmol/l และ โมล/l ดูไอออนตั้งโปรแกรม P3.0 (หน้า 43) เพื่อเลือกหน่วยความเข้มข้นของไอออนเทียบไอออนเครื่องมือนี้มีความสามารถในถึงไอออนเอง 5 จุดเทียบ ด้วยอย่างน้อย 2 จุดเทียบ ค่าปรับเทียบแบบกำหนดเองสามารถมีค่าระหว่าง 0.001 และ 19998 เมื่อเทียบค่าแรกคือ keyed ใน เทียบอื่น ๆ จะมีน้อยหนึ่งทศวรรษที่แยกโดยอัตโนมัติ สำหรับตัวอย่างถ้าคุณเทียบแรกชี้อยู่ที่ 0.001 เทียบจุดอื่น ๆ จะน้อยหนึ่งทศวรรษจาก 0.001 (0.001, 0.01, 0.1,1.0, 10.0, 100, 1000, 10000) สำหรับความแม่นยำสุดปรับเทียบเครื่องวัดจุดกับความเข้มข้นคล้ายกับโซลูชั่นคุณต้องการทดสอบ 1. 2 เชื่อมต่อเป็นไอออนใช้อิเล็กโทรด (อิเสะ) ขั้วต่อ BNC เข้าบนด้านหลังของตัววัด เปิดมิเตอร์ ถ้ามิเตอร์ไม่ได้อยู่ในโหมดวัดไอออน กดออก และใช้ > / CAL เพื่อเลือกไอออน โหมดที่เลือกที่จะแฟลช ดูรูป 48 กด enter เพื่อยืนยันการเลือก ตัววัดกำลัง m วัดไอออน
การแปล กรุณารอสักครู่..

PREFACEPREFACE
Thank you for selecting the pH 1100/ 2100 benchtop meter. This meter measure pH, millivolts, relative millivolts and temperature. The pH 2100 meter also measures ion concentration. The instruction manual serves to explain the use of the pH 1100/ 2100 bench meter as a step-by-step operational guide to help you familiarize with the meters features and functions. It is structured sequentially with illustration of diagrams that explains the various functions and setup menus available. This manual is written to cover as many anticipated applications and uses of the pH 1100/ 2100 Bench meter as possible. If there are doubts in the use of the meter, please do not hesitate to contact the nearest Authorized Distributor or Customer Service Dept. for assistance. Eutech Instruments/ Oakton Instruments reserve the rights to change, make improvement and modify specifications without prior notice and cannot accept any responsibility for damage or malfunction to the instrument caused by improper use.
Copyright 2002 Eutech Instruments Pte. Ltd. Oakton Instruments All rights reserved. Revision 2.2, Nov 02.
TABLE OF CONTENTS
1 INTRODUCTION 1.1 1.2 1.3 1.Introducing the Bench Meter Series Keypad Rear Panel Electrode Holder 35
STARTING UP 2.1 Back Panel Connections 2.2 Powering up and powering down 2.2.1 Powering up 2.2.2 Powering down
CALIBRATION & MEASUREMENT
3.1 pH calibration 3.1.1 Automatic Temperature Compensation (ATC) 3.1.2 Starting pH calibration 3.1.3 Standard pH buffer calibration 3.1.4 Custom pH buffer calibration 3.1.5 Calibration error messages 3.2 pH measurement 3.2.1 Automatic temperature compensation 3.2.2 Adjusting manual temperature compensation 3.2.3 Taking pH Measurements 4 mV CALIBRATION AND MEASUREMENT 4.1 mV Calibration 4.1.1 mV Calibration error message 4.2 mV Measurement 5 ION CALIBRATION AND MEASUREMENT (ONLY APPLICABLE FOR pH 2100) 5.1 5.2 5.Ion calibration Ion calibration error messages Ion measurement
MEMORY FUNCTIONS 6.1 Memory input 6.1.1 Manually storing a reading into memory 6.2 Memory recall 6.2.1 Recalling readings in manual recall mode 6.2.2 Recalling readings in automatic recall mode
STABILITY INDICATOR ALARM FUNCTIONS 8.1 8.2 High and Low measurement alarm Calibration due alarm Temperature Calibration Temperature Calibration Error Messages
TEMPERATURE CALIBRATION 9.1 9.2
10 SETUP MODE 10.1 Setup mode overview 10.1.1 pH/Temperature setup submenus 10.1.2 mV setup submenus 10.1.3 Ion setup submenus (pH 2100 only) 10.1.4 Meter general configuration setup submenus 10.2 pH/ temperature setup mode 10.2.1 Entering pH/temperature setup mode 10.2.2 pH buffer setup program P1.0 10.2.3 pH resolution setup program P1.1 10.2.4 Temperature unit setup program P1.2 10.2.5 pH Measurement Alarm Setup Program P1.3 10.2.6 pH Calibration Due Alarm Setup program P1.4 10.2.7 Stability Setup Program P1.5 10.2.8 View pH calibration data setup program P1.6
10.2.9 pH calibration data reset setup program P1.7 10.3 mV setup mode 10.3.1 Entering mV setup mode 10.3.2 mV measurement alarm setup program P2.0 10.3.3 mV calibration due alarm setup program P2.1 10.3.4 Stability setup program P2.2 10.3.5 View mV calibration data setup program P2.3 10.3.6 mV calibration data reset setup program P2.4 10.4 Ion setup mode (pH 2100 meter only) 10.4.1 Entering ion setup mode 10.4.2 Ion unit setup mode P3.0 10.4.3 Ion measurement alarm setup program P3.1 10.4.4 Ion calibration due alarm setup program P3.2 10.4.5 Stability setup program P3.3 10.4.6 Ion mode setup program P3.4 10.4.7 View ion calibration data setup program P3.5 10.4.8 Ion calibration data reset setup program P3.6 10.5 Meter general configuration setup mode 10.5.1 Entering meter general setup mode 10.5.2 Date/time setup mode P4.0 10.5.3 Backlight setup mode P4.1 10.5.4 Data log setup mode P4.2 10.5.5 Memory recall setup mode P4.3 10.5.6 Communication setup mode P4.4 10.5.7 Data transfer setup mode P4.5 10.5.8 Memory clear setup mode P4.6 10.5.9 Meter reset setup mode P4.ELECTRODE CARE 11.1 11.2 11.3 Electrode Activation Electrode Maintenance Storing pH/ORP electrodes
Connector
ATC INPUT
Function
For phono jack connection from the temperature probe for Automatic Temperature Compensation. The probe should be a 30K thermistor probe. For connecting sensors with a BNC connector to the meter. The meter accepts any pH, ORP or ISE electrode with a BNC connector. Always make sure that the connector is clean and dry. For connection to the pin tip type reference electrode normally used with half cell (mono) type pH or ISE electrodes. For connection to the strip chart recorder. For connection to the ground earth jack (standard pin tip connectors). For connection to the RS232C serial port. For connection of the AC adapter power supply (included).
REF REC GND RS232 DC
Electrode Holder
The electrode holder is included in the same box as the meter.
To attach the electrode holder to the meter: 1. The electrode holder base attached to the bottom of the meter comes in the shipping position. See Figure 3. Use a Phillips screwdriver to remove the screw holding the electrode holder in shipping position. See Figure 3. Slide the electrode base away from the meter until the second screw slot lines up with the original screw hole. Use the screw removed earlier to secure electrode holder base into position. See Figure 3. The electrode holder arm is reversible. If desired, remove screw holding electrode holder base and slide base out of brackets. Slide base into brackets on opposite side and replace screw. To install the electrode arm, turn meter back to the upright position. See Figure 4. Line up the hole on the base of the electrode holder arm with the peg on the electrode holder base. Slide the hole securely onto the electrode holder base. See Figure 4. The electrode arm is now ready to swing into desired position.
Figure 3: Bottom of Meter
Figure 4: Fix electrode holder arm to holder base
STARTING UP
Attention: Do not get water on the BNC connector during operation. Avoid touching the connector with soiled or wet hands.
2.1 Back Panel Connections
Refer back to Figure 2. Make the necessary connections as mentioned in Section 1.3 Rear Panel, pg 8.
2.2.1 1.
Powering up and powering down
3.1.3 1.
Standard pH buffer calibration
When entering the standard calibration mode, the first buffer calibration point for standard pH buffers is 7.000 or 6.860 (depending upon which pH buffer set is selected). The upper display shows the current reading, the lower display will show Auto Buf Scan: 7.000 (or 6.860). The upper display will flash pH until the reading has stabilized. See Figure 16. If the selected buffer value is not within 1.0 of the measured pH value, the Auto pH Scan will continue to change until the pH buffer is within 1.0. When pH reading has stabilized, press ENTER. The lower display flashes *pH Offset Cal Done. See Figure 17. This calibration point is now stored in the meter. If you are performing a one-point calibration, press EXIT. The screen flashes Sys Updating.Please wait.The meter then returns to the pH measurement mode. You may now start taking pH readings. If you are performing a multi-point calibration, go to step 3.
7.001pH 25.4C ATC Auto Buf Scan: 7.000
Figure 16
7.001pH 25.4C ATC *pH Offset Cal Done
Figure 17
ATC 10.010pH 25.4C 2 Point Cal Done *
Figure 18
Rinse the probe with deionized water or a rinse solution, and place it in the next pH buffer. The lower display automatically scans to the next pH buffer solution. Wait for the measured pH value to stabilize. The upper display will flash pH until the reading has stabilized. When pH reading has stabilized, press ENTER. The lower display flashes 2 Point Cal Done * (3 if you have calibrated 3 points,4 if 4 points, etc.). See Figure 18. This calibration point is now stored in the meter. Repeat steps 3-5 for each point you wish to calibrate. When you are finished calibrating, press EXIT. The screen flashes Sys. Updating. Please wait. The meter then returns to the pH measurement mode. You may now start taking pH readings. If you perform a complete 5-point pH calibration, when you calibrate the fifth point and press ENTER, the meter automatically returns to measurement mode. You may now start taking pH readings.
Error: Cal Aborted
Figure 45
The meter is now in Rel mV measurement mode. See Figure 46. If you have not performed a mV calibration or have reset the meter, the display will read Abs. mV. See Figure 47. 2. Dip the electrode into the sample. The sensor or glass bulb of the electrode must be completely immersed into the sample. Stir the electrode gently to create a homogeneous sample. Allow time for the reading to stabilize. The upper display will show the mV reading, and the lower display will show the current month, day and time. See Figure 46.
Figure 46
Figure 47
ION CALIBRATION AND MEASUREMENT (ONLY APPLICABLE FOR pH 2100)
The pH 2100 meter reads ion concentration in ppm, ppt, mg/l, g/l, mmol/l and mol/l units. See Ion setup Program P3.0 (page 43) to select ion concentration units.
Ion calibration
This instrument is capable of up to 5-point custom ion calibration with a minimum of 2-point calibration. Custom calibration values can be any value between 0.001 and 19998. Once the first calibration value is keyed in, all other calibration will automatically be at least one decade apart. For example if your first calibration point is at 0.001, all other calibration points will be at least one decade apart from 0.001 (0.001, 0.01, 0.1,1.0, 10.0, 100, 1000, 10000). For best accuracy calibrate your meter to points with similar concentrations to the solutions you want to test. 1. 2. Connect an Ion Selective Electrode (ISE) to the BNC input connector on the back of the meter. Turn the meter ON. If the meter is not in the Ion measurement mode, press EXIT and use >/CAL to select Ion. The mode selected will flash. See Figure 48. Press ENTER to confirm selection. The meter is now in Ion measurement m
การแปล กรุณารอสักครู่..

prefacepreface
ขอบคุณสำหรับการเลือก M 1100 / 2100 เมตร แบบตั้งโต๊ะ . นี้มิเตอร์วัด pH มิลลิโวลต์มิลลิโวลต์ญาติ , และอุณหภูมิ เครื่องวัด pH 2100 ยังวัดความเข้มข้นของไอออน . คู่มือการใช้บริการเพื่ออธิบายการใช้ pH 1100 / 2100 เมตรเป็นคู่มือทีละขั้นตอนม้านั่งปฏิบัติการที่จะช่วยให้คุณคุ้นเคยกับเมตรคุณสมบัติและฟังก์ชั่นมันมีโครงสร้างเป็นภาพประกอบของไดอะแกรมอธิบายฟังก์ชั่นต่างๆและเมนูการติดตั้งพร้อมใช้งาน คู่มือนี้เขียนครอบคลุมหลายคาดการใช้งานและการใช้เครื่องวัด pH 1100 / 2100 ม้านั่งที่เป็นไปได้ หากมีข้อสงสัยในการใช้เครื่องวัด โปรดไม่ลังเลที่จะติดต่อที่ใกล้ที่สุดที่ได้รับอนุญาตจำหน่ายหรือฝ่ายบริการลูกค้าเพื่อขอความช่วยเหลืออุณหภูมิ เครื่องมือ / อุปกรณ์ oakton สงวนสิทธิที่จะเปลี่ยนแปลง ให้ปรับปรุงและแก้ไขข้อมูลโดยไม่ต้องแจ้งให้ทราบล่วงหน้า และไม่รับผิดชอบต่อความเสียหาย หรือ ผิดปกติกับเครื่องมือที่เกิดจากการใช้ที่ไม่เหมาะสม
ลิขสิทธิ์ 2002 จำกัด oakton อุณหภูมิเครื่องมือเครื่องมือสงวนสิทธิ์ทั้งหมด แก้ไข 2.2 พ.ย. 02
สารบัญ
1 บทนำ 1.1 1.2 1.3 1แนะนำชุดม้านั่งกด มิเตอร์ไฟฟ้ายึดแผงหลัง 35
เริ่มต้น 2.1 2.2 เปิดการเชื่อมต่อแผงหลังและลงโปรแกรมเปิดเปิดเปิดลง 2.2.2 สอบเทียบการวัด pH &
3.1 การปรับชดเชยอุณหภูมิอัตโนมัติ ( ATC ) 3.1.1 3.1.2 3.1.3 pH buffer pH เริ่มต้นการสอบเทียบมาตรฐานการสอบเทียบ pH buffer สอบเทียบเอง 3.1.4 3.1 .5 ปรับแต่งข้อความแสดงข้อผิดพลาด 3.2 เครื่องวัดดำเนินงานมีระบบชดเชยอุณหภูมิอัตโนมัติ 3.2.2 ปรับชดเชยอุณหภูมิ 3.2.3 พาอคู่มือการวัดและสอบเทียบการวัด 4 MV MV เพลง calibration สอบเทียบ 4.1 4.1.1 ข้อผิดพลาด 4.2 การวัดและสอบเทียบการวัดไอออน MV 5 ( เฉพาะใช้ได้กับ pH 2100 ) 5.1 5.2 5ไอออนไอออนการสอบเทียบข้อผิดพลาดการทำงาน 6.1 หน่วยความจำและการจัดเก็บข้อมูลด้วยตนเองอ่านในหน่วยความจำหน่วยความจำเรียกคืนเรียกคืน 6.2.1 6.2 อ่านในคู่มือว่าโหมด 6.2.2 ระลึกถึงอ่านในโหมดอัตโนมัติเรียกคืนเสถียรภาพบ่งชี้เตือนฟังก์ชัน 8.1
8 หน่วยการวัดไอออน2 สูงต่ำและการวัดอุณหภูมิเนื่องจากอุณหภูมิปลุกปลุกสอบเทียบสอบเทียบสอบเทียบสอบเทียบอุณหภูมิ 9.1 9.2 ข้อความผิดพลาด
10 การตั้งค่าโหมด 10.1 ติดตั้งโหมดภาพรวม 10.1.1 pH / อุณหภูมิการตั้งค่าเมนูย่อย 10.1.2 MV ติดตั้งเมนูย่อย 10.1.3 ไอออนติดตั้งเมนูย่อย ( pH 2100 เท่านั้น ) 10.1.4 เครื่องวัดค่า pH และอุณหภูมิน้ำทั่วไปการตั้งค่าเมนูย่อยการตั้งค่าโหมด 10.2 .1 การติดตั้งโหมด 10.2.2 pH อุณหภูมิ pH buffer pH ละเอียดโปรแกรมติดตั้ง p1.0 10.2.3 โปรแกรมติดตั้ง p1.1 10.2.4 หน่วยอุณหภูมิติดตั้งโปรแกรม p1.2 10.2.5 เครื่องวัดสัญญาณติดตั้งโปรแกรม p1.3 10.2.6 Ph สอบเทียบเนื่องจากโปรแกรมติดตั้ง p1.4 10.2.7 เสถียรภาพปลุกโปรแกรมติดตั้ง p1.5 10.2.8 ดูการติดตั้งโปรแกรมอข้อมูล p1.6
10.2.9 สอบเทียบ pH รีเซ็ตข้อมูลการติดตั้งโปรแกรม p1.7 10การตั้งค่าโหมดการตั้งค่าโหมด 10.3.1 3 MV MV MV ปลุก 10.3.2 การวัดโปรแกรมติดตั้ง p2.0 10.3.3 MV สอบเทียบจากโปรแกรมติดตั้ง p2.1 10.3.4 เสถียรภาพปลุกโปรแกรมติดตั้ง p2.2 10.3.5 ดู MV สอบเทียบข้อมูลโปรแกรม p2.3 10.3.6 MV สอบเทียบข้อมูลใหม่การติดตั้งโปรแกรมการติดตั้งโหมด p2.4 10.4 ไอออน ( pH 2100 เมตรเท่านั้น ) 10.4.1 ป้อนการตั้งค่าโหมด 10.4.2 หน่วยติดตั้งไอออนไอออน โหมด p3.0 10.4 .3 รายละเอียดการวัดสัญญาณเตือนภัยติดตั้งโปรแกรม p3.1 10.4.4 ไอออนการสอบเทียบจากโปรแกรมติดตั้ง p3.2 10.4.5 เสถียรภาพปลุกโปรแกรมติดตั้ง p3.3 10.4.6 ไอออนโหมดโปรแกรม p3.4 10.4.7 ดูข้อมูลการติดตั้งโปรแกรม p3.5 ไอออนไอออน 10.4.8 สอบเทียบข้อมูลตั้งค่าโปรแกรม p3.6 10.5 เมตร การตั้งค่าโหมดการตั้งค่าทั่วไป 10.5.1 เข้าโหมด 10.5.2 เครื่องวัดทั่วไปการตั้งค่าวัน / เวลาการตั้งค่าโหมด p4.0 10.53 การตั้งค่าโหมด Backlight p4.1 10.5.4 ข้อมูลเข้าสู่ระบบการตั้งค่าโหมด p4.2 10.5.5 หน่วยความจำเรียกคืนการตั้งค่าโหมด p4.3 10.5.6 การสื่อสารตั้งค่าโหมดการถ่ายโอนข้อมูล p4.4 10.5.7 ตั้งค่าโหมด p4.5 10.5.8 ล้างความจำการตั้งค่ารีเซ็ตการตั้งค่าโหมดโหมด p4.6 10.5.9 เครื่องวัด p4.electrode ดูแลด้านการซ่อมบำรุงไฟฟ้าจำนวน 11.3 จัดเก็บ pH / ORP อิเล็กโทรดเชื่อมต่อขั้วไฟฟ้า
ATC ใส่ฟังก์ชั่น
สำหรับเสียงแจ็คการเชื่อมต่อจากอุณหภูมิโพรบสำหรับการชดเชยอุณหภูมิอัตโนมัติ การสอบสวนน่าจะ 30K Thermistor Probe . สำหรับการเชื่อมต่อเซ็นเซอร์กับ BNC Connector ที่จะเมตร เมตรยินดีใด ๆหรือ pH ORP อิเล็กโทรอิเสะกับ BNC Connector เสมอให้แน่ใจว่าการเชื่อมต่อมีความสะอาด และแห้งสำหรับการเชื่อมต่อกับพินทิปประเภทอ้างอิงขั้วปกติใช้กับครึ่งเซลล์ ( Mono ) ประเภท : pH หรือขั้วไฟฟ้า สำหรับการเชื่อมต่อกับแถบบันทึกแผนภูมิ . สำหรับการเชื่อมต่อกับพื้นดิน แจ็ค ( ปลายขามาตรฐานการเชื่อมต่อ ) สำหรับการเชื่อมต่อกับ rs232c พอร์ตอนุกรม สำหรับการเชื่อมต่อของอะแดปเตอร์ไฟฟ้า ( รวม )
Ref rec GND RS232 DC ขั้วยึด
ขั้วใส่รวมในกล่องเดียวกันเป็นเครื่องวัด .
ติดขั้วไฟฟ้าที่ยึดในการเมตร : 1 ขั้วยึดฐานติดกับด้านล่างของมิเตอร์มาส่งตำแหน่ง ดูรูปที่ 3 ใช้ไขควงแบน เอาสกรูยึดถือขั้วตำแหน่ง Shipping ดูรูปที่ 3สไลด์ไฟฟ้าฐานห่างจากวัดจนถึงช่องสกรูสองเส้นกับหลุมสกรูเดิม ใช้สกรูลบออกก่อนหน้านี้เพื่อรักษาความปลอดภัยฐานยึดขั้วไฟฟ้าลงในตำแหน่ง ดูรูปที่ 3 ขั้วยึดแขนผันกลับได้ . ถ้าต้องการเอาสกรูยึดถือเบสไฟฟ้าและฐาน ภาพนิ่งจากวงเล็บ เลื่อนฐานเป็นวงเล็บบนด้านตรงข้ามและแทนที่สกรูการติดตั้งไฟฟ้าที่แขนเปิดเมตรกลับตำแหน่งตรง เห็นรูปที่ 4 บรรทัดขึ้นหลุมบนฐานของขั้วยึดแขนกับหมุดบนขั้วยึดฐาน สไลด์รูอย่างปลอดภัยบนขั้วยึดฐาน เห็นรูปที่ 4 อิเล็กโทรดที่แขนคือตอนนี้พร้อมที่จะแกว่งลงไปในตำแหน่งที่ต้องการ ด้านล่างของรูปที่ 3 :
รูปที่ 4 : แก้ไขมิเตอร์ไฟฟ้าเพื่อยึดแขนยึดฐาน
เริ่มต้นขึ้น
สนใจ : ไม่ได้รับน้ำบนขั้วต่อบีเอ็นซีในระหว่างการผ่าตัด หลีกเลี่ยงการสัมผัส หรือการเชื่อมต่อกับเปื้อนมือเปียก .
2.1 แผงหลังการเชื่อมต่อ
กลับมาดูรูปที่ 2 ให้การเชื่อมต่อที่จำเป็นตามที่ระบุไว้ในมาตรา 1.3 แผงด้านหลัง , PG 8 .
ร 1 .
เปิดเครื่องขึ้นเครื่องลง 3.1.3 และ
1 . มาตรฐานการสอบเทียบพีเอชบัพเฟอร์
เมื่อป้อนสารมาตรฐานในโหมดบัฟเฟอร์การสอบเทียบจุดแรกสำหรับบัฟเฟอร์ pH 7 หรือมาตรฐานคือ 6.860 ( ซึ่งขึ้นอยู่กับ pH Buffer ตั้งค่าเลือก ) จอภาพด้านบนแสดงการอ่านปัจจุบัน จอล่างจะแสดงรถยนต์ buf สแกน : 7.000 ( หรือ 6.860 ) จอแสดงผลบนจะแฟลชอ จนอ่านได้คงที่ รูปที่ 16 ถ้าเลือกบัฟเฟอร์มูลค่าไม่เกิน 1.0 ของวัด pH ค่าออัตโนมัติสแกนจะเปลี่ยนไปเรื่อย จนถึง pH buffer ภายใน 1.0 เมื่อ pH อ่านมีเสถียรภาพ กด Enter กะพริบ * แสดงลด pH ชดเชยแคลเสร็จแล้ว รูปที่ 17 ตอนนี้เก็บไว้ในจุดนี้การสอบเทียบมาตรวัด ถ้าคุณจะแสดงเป็นจุดหนึ่งการสอบเทียบ , กดออก กะพริบหน้าจอระบบปรับปรุง โปรดรอ มิเตอร์แล้วส่งกลับไปยังเครื่องวัดโหมดตอนนี้คุณอาจเริ่มการอ่านค่า PH ถ้าคุณมีการปรับแต่งหลายจุด ไปที่ขั้นตอนที่ 3
7.001ph 25.4c ATC Auto buf สแกน : 7.000
รูปที่ 16
7.001ph 25.4c ATC * Ph ชดเชยแคลทำ
รูปที่ 17
ATC 10.010ph 25.4c 2 จุดแคลทำ *
รูปที่ 18
ล้างตัวด้วยน้ำคล้ายเนื้อเยื่อประสานหรือ ล้างออกโซลูชันและวางไว้ในบัฟเฟอร์ pH ต่อไปจอแสดงผลลดลงโดยอัตโนมัติสแกนไปยัง pH สารละลายบัฟเฟอร์ รอวัดค่า pH ให้คงที่ จอแสดงผลบนจะแฟลชอ จนอ่านได้คงที่ เมื่อ pH อ่านมีเสถียรภาพ กด Enter แสดงราคาเป็น 2 จุด แคลา * ( ถ้าคุณได้ปรับจุด 3 , 4 ถ้า 4 จุด , ฯลฯ ) รูปที่ 18 ตอนนี้เก็บไว้ในจุดนี้การสอบเทียบมาตรวัดขั้นตอนที่ 5 ทำซ้ำสำหรับแต่ละจุดที่คุณต้องการที่จะปรับ . เมื่อคุณเสร็จสิ้นการปรับเทียบ กดออก ที่หน้าจอกระพริบ SYS . การปรับปรุง โปรดรอสักครู่ เครื่องวัดเครื่องวัดแล้วส่งกลับไปยังโหมด ตอนนี้คุณอาจเริ่มการอ่านค่า PH ถ้าคุณดำเนินการเสร็จสมบูรณ์และการสอบเทียบ pH เมื่อคุณปรับจุดที่ห้าและกดมิเตอร์โดยอัตโนมัติกลับไปที่โหมดการวัดตอนนี้คุณอาจเริ่มการอ่านค่า pH .
ผิดพลาด : แคลแล้ว
รูปที่ 45 เมตร ตอนนี้อยู่ในโหมดการวัด rel MV . ดูรูปที่ 46 ถ้าคุณไม่ได้ทำการสอบเทียบเครื่องวัดหรือ MV ได้ใหม่ จอจะอ่านจาก MV . ดูรูปที่ 47 2 . จุ่มขั้วไฟฟ้าเข้าไปในตัวอย่าง เซ็นเซอร์หรือกระจก หลอดไฟของขั้วไฟฟ้าต้องแช่อย่างสมบูรณ์ในตัวอย่างผัดไฟฟ้าเบา ๆเพื่อสร้างอย่างเป็นเนื้อเดียวกัน ให้เวลาในการอ่านให้คงที่ จอแสดงผลบนจะแสดง MV อ่านและแสดงที่ด้านล่างจะแสดงเดือนปัจจุบัน , วันที่และเวลา ดูรูปที่ 46 .
รูปที่ 2
รูปที่ 47 และสอบเทียบการวัดไอออน ( แต่ใช้ได้กับ pH 2100 )
pH 2100 เมตรอ่านความเข้มข้นของไอออนใน ppm ( mg / L , G / L , หน่วย mmol / L และ mol / L .ดูรายละเอียดการติดตั้งโปรแกรม p3.0 ( หน้า 43 ) เพื่อเลือกหน่วยความเข้มข้นไอออนไอออนค่า
นี้เครื่องมือที่สามารถขึ้นงานเองรายละเอียดการสอบเทียบกับต่ำสุดของเช่นสอบเทียบ การปรับค่าเองสามารถเป็นค่าใด ๆระหว่าง 0.001 และ 19998 . เมื่อค่าแรกคือ คีย์ในการสอบเทียบ , สอบเทียบทั้งหมดอื่น ๆโดยอัตโนมัติจะได้รับอย่างน้อยหนึ่งทศวรรษ ต่างหากตัวอย่างเช่นถ้าจุดสอบเทียบแรกของคุณคือที่ 0.001 , อื่น ๆทั้งหมดที่สอบเทียบจุดจะมีอย่างน้อยหนึ่งทศวรรษนอกจาก 0.001 ( 0.001 , 0.01 , 0.1,1.0 , 10.0 , 100 , 1000 , 10000 ) ที่ดีที่สุดสำหรับการสอบเทียบเครื่องวัดความของคุณไปยังจุดที่มีความเข้มข้นใกล้เคียงกับโซลูชั่นที่คุณต้องการทดสอบ 1 . 2 . เชื่อมต่อไอออน selective electrode ( ISE ) กับขั้วต่อ BNC input ด้านหลังของเมตรเปิดเมตร . ถ้าวัดไม่ได้อยู่ในโหมดการวัดไอออนออกกดใช้ > / แคลเลือกไอออน โหมดที่เลือกจะแฟลช ดูรูปที่ 48 กด Enter เพื่อยืนยันการเลือก เมตรอยู่ใน M วัดไอออน
การแปล กรุณารอสักครู่..