Quantitative investigation of a standard test shim for magnetic particle testing has been carried out. The current values of the magnetizing yoke to obtain magnetic particle indication were measured experimentally. An analytical model for the experimental conditions was then formulated to obtain the amplitude of the leakage magnetic flux from an artificial groove in a type A standard test shim. The magnetic flux densities in the type A standard test shim and the test object were obtained using finite element analysis.