A ‘cold chain breach’ has occurred if vaccine storage temperatures have been
outside the recommended range of +2°C to +8°C. It excludes fluctuations up to
+12°C, lasting no longer than 15 minutes, as may occur when stock taking or
restocking refrigerators.