5. Sampling
5.1 Choice of sampling point
Discharge sample points must be at a location that ensures that the sample is truly
representative of the discharge.
The sampling point location should be agreed with us, documented and clearly and
permanently labelled.
A sampling position in a pipe or channel must be sufficiently far downstream of the last
inflow that mixing of the two streams is complete.
Samples at an outfall should be taken from regions of high turbulence and good mixing,
usually at the centre of the discharge. Solid materials will have little chance to settle out
here.
Samples in channels should be collected away from the sides and bottom of the channel
to avoid contamination of the sample with sediment and biological growths.
When sampling from chambers (for example manholes), avoid contamination of the
sample by the disturbance of deposits from the cover when the cover is lifted and prevent
contamination of the sample from the chamber walls and any bottom deposits.
Samples may also be drawn off from effluent streams at a tap. Care should be taken to
ensure any dead space is flushed out with effluent before the sample is collected.
Sampling staff should be aware that manholes and similar confined spaces are
dangerous and must not be entered unless in accordance with a safe system of work
and after appropriate training.
When automatic samplers are employed for composite sampling care should be taken to
ensure the sample probe is deployed taking the above factors into consideration. Also it is
important to ensure that the probe remains in the effluent flow during the entire period each
sample aliquot is being taken i.e. variations in effluent flow should not result in the sample
probe being left dangling in the air or in contact with the bottom of the channel.
Effluent may be discharged in batches from hold-up tanks, for example, discharge may only
be permitted at high tides. Wherever possible, the tanks should be mixed, and in some cases,
samples are taken from recirculation lines. If good mixing is not possible, it may be necessary
to increase the frequency of sampling during discharge.
5.2 Composite or spot samples
There are two main sampling methods identified for wastewater, composite sampling and spot
sampling:
Composite samples. Two types of composite sample are commonly used, flow-proportional
and time-proportional. For a flow-proportional sample, a fixed amount of sample is taken for
each pre-defined volume of effluent (for example every 10 m3
). For time-proportional samples,
a fixed amount of sample is taken from the effluent for each time unit (for example every
M18: Monitoring of discharges to water and sewer
M18 Version 4 November 2014 Page 9 of 48
hour). The analysis of a composite sample gives an average value of the determinand during
the period over which the sample has been collected. It is normal to collect composite
samples over 24 hours to give a daily mean value. Shorter times can be used with prior
agreement from us.
Flow proportional sampling should be employed when the volume of effluent discharged
varies significantly throughout the sampling period. When the volume of discharge is relatively
constant then time proportional composites are appropriate.
Automatic sampling equipment is recommended (see section 5.3). However, it is necessary to
consider the stability of the target substances over the total sample collection time, as
samples may deteriorate while sitting in the automatic sampling device.
Spot samples. These are discrete samples taken at random time intervals from a discharge,
and are not related to volume of discharge. They are best suited where:
the composition of the waste water is relatively constant
the discharge contains mineral oil or volatile substances, or when, due to decomposition,
evaporation or coagulation, the target substances are not stable in the sample
separate phases are present (for example an oil layer floating on water)
there is a need to check the quality of the discharge at a particular moment, normally to
assess compliance with the permit conditions
the discharge is not continuous (from batch or hold-up tanks), but only when the effluent is
well mixed
collecting larger object and floating matter that is not representative of the discharge may
be possible.
Bulked composites. When a bulk sample is prepared by manually compositing a series of
samples in a laboratory, it is important to consider the stability of the determinand being
measured. For example, BOD will start to deteriorate significantly after 24 hours.
It may not be appropriate to collect composite samples for periods greater than 24 hours, due
to the stability of some determinands, even when autosamplers are refrigerated. For example,
BOD, pH, COD and ammonia.
5.3 Automatic sampling equipment and MCERTS
We require that automatic sampling devices used for self-monitoring purposes have been
tested and certified to the MCERTS performance standard:
‘Conti
5. Sampling5.1 Choice of sampling pointDischarge sample points must be at a location that ensures that the sample is trulyrepresentative of the discharge. The sampling point location should be agreed with us, documented and clearly andpermanently labelled. A sampling position in a pipe or channel must be sufficiently far downstream of the lastinflow that mixing of the two streams is complete. Samples at an outfall should be taken from regions of high turbulence and good mixing,usually at the centre of the discharge. Solid materials will have little chance to settle outhere. Samples in channels should be collected away from the sides and bottom of the channelto avoid contamination of the sample with sediment and biological growths. When sampling from chambers (for example manholes), avoid contamination of thesample by the disturbance of deposits from the cover when the cover is lifted and preventcontamination of the sample from the chamber walls and any bottom deposits. Samples may also be drawn off from effluent streams at a tap. Care should be taken toensure any dead space is flushed out with effluent before the sample is collected.Sampling staff should be aware that manholes and similar confined spaces aredangerous and must not be entered unless in accordance with a safe system of workand after appropriate training.When automatic samplers are employed for composite sampling care should be taken toensure the sample probe is deployed taking the above factors into consideration. Also it isimportant to ensure that the probe remains in the effluent flow during the entire period eachsample aliquot is being taken i.e. variations in effluent flow should not result in the sampleprobe being left dangling in the air or in contact with the bottom of the channel.Effluent may be discharged in batches from hold-up tanks, for example, discharge may onlybe permitted at high tides. Wherever possible, the tanks should be mixed, and in some cases,samples are taken from recirculation lines. If good mixing is not possible, it may be necessaryto increase the frequency of sampling during discharge.5.2 Composite or spot samplesThere are two main sampling methods identified for wastewater, composite sampling and spotsampling:Composite samples. Two types of composite sample are commonly used, flow-proportionaland time-proportional. For a flow-proportional sample, a fixed amount of sample is taken foreach pre-defined volume of effluent (for example every 10 m3). For time-proportional samples,a fixed amount of sample is taken from the effluent for each time unit (for example every M18: Monitoring of discharges to water and sewerM18 Version 4 November 2014 Page 9 of 48hour). The analysis of a composite sample gives an average value of the determinand duringthe period over which the sample has been collected. It is normal to collect compositesamples over 24 hours to give a daily mean value. Shorter times can be used with prior
agreement from us.
Flow proportional sampling should be employed when the volume of effluent discharged
varies significantly throughout the sampling period. When the volume of discharge is relatively
constant then time proportional composites are appropriate.
Automatic sampling equipment is recommended (see section 5.3). However, it is necessary to
consider the stability of the target substances over the total sample collection time, as
samples may deteriorate while sitting in the automatic sampling device.
Spot samples. These are discrete samples taken at random time intervals from a discharge,
and are not related to volume of discharge. They are best suited where:
the composition of the waste water is relatively constant
the discharge contains mineral oil or volatile substances, or when, due to decomposition,
evaporation or coagulation, the target substances are not stable in the sample
separate phases are present (for example an oil layer floating on water)
there is a need to check the quality of the discharge at a particular moment, normally to
assess compliance with the permit conditions
the discharge is not continuous (from batch or hold-up tanks), but only when the effluent is
well mixed
collecting larger object and floating matter that is not representative of the discharge may
be possible.
Bulked composites. When a bulk sample is prepared by manually compositing a series of
samples in a laboratory, it is important to consider the stability of the determinand being
measured. For example, BOD will start to deteriorate significantly after 24 hours.
It may not be appropriate to collect composite samples for periods greater than 24 hours, due
to the stability of some determinands, even when autosamplers are refrigerated. For example,
BOD, pH, COD and ammonia.
5.3 Automatic sampling equipment and MCERTS
We require that automatic sampling devices used for self-monitoring purposes have been
tested and certified to the MCERTS performance standard:
‘Conti
การแปล กรุณารอสักครู่..

5 . การสุ่มตัวอย่าง5.1 เลือกสุ่มจุดคะแนนตัวอย่างโดยไม่ต้องมีในสถานที่ที่ช่วยให้มั่นใจว่า ตัวอย่างจริงๆตัวแทนของการปล่อยการสุ่มตัวอย่างจุดที่ตั้งควรตกลงกับเราไว้แล้วอย่างชัดเจน และถาวรข้อความสุ่มตำแหน่งในท่อหรือช่องต้องเพียงพอไกลด้านสุดท้ายกระแสที่ผสมสองสายสมบูรณ์ตัวอย่างที่ outfall ควรถ่ายจากภูมิภาคปั่นป่วนสูงและดีผสมปกติจะอยู่ที่ศูนย์กลางของการปล่อย วัสดุของแข็งจะมีโอกาสน้อยที่จะจ่ายออกที่นี่ตัวอย่างในช่องทางที่ควรเก็บห่างจากด้านข้างและด้านล่างของรางเพื่อหลีกเลี่ยงการปนเปื้อนของตัวอย่างดินตะกอน และการเจริญเติบโตทางชีวภาพเมื่อจำนวนห้อง ( ตัวอย่างเช่น manholes ) หลีกเลี่ยงการปนเปื้อนของตัวอย่าง โดยการรบกวนของฝากจากปกเมื่อฝาจะถูกยกขึ้นและป้องกันการปนเปื้อนของเชื้อโรค จากผนังห้อง และฐานเงินฝากตัวอย่างยังอาจจะถูกดึงออกจากธารน้ำที่เคาะ ควรดูแลให้ให้แน่ใจว่าพื้นที่ใดตายจะขับออกมากับน้ำทิ้งก่อน ตัวอย่างจะถูกเก็บรวบรวมพนักงานตัวอย่างควรทราบว่า manholes และที่คล้ายกันคับเป็นเป็นอันตราย และไม่ต้องใส่ ยกเว้นตามระบบเซฟงานและหลังการฝึกอบรมที่เหมาะสมเมื่อตัวอย่างอัตโนมัติ ในการสุ่มตัวอย่างควรที่จะดูแลคอมโพสิตให้ตัวอย่างตัวจัดการปัจจัยข้างต้นมาพิจารณา นอกจากนี้ที่สำคัญเพื่อให้แน่ใจว่า การสอบสวนยังคงไหลออกตลอดช่วงระยะเวลาแต่ละตัวอย่าง เช่น การเปลี่ยนแปลงในการเทศนาคือถูกกระแสน้ำจะไม่ส่งผลในตัวอย่างสอบสวน ถูกทิ้งให้ห้อยอยู่ในอากาศ หรือในการติดต่อกับด้านล่างของช่องน้ำอาจจะออกจากโรงพยาบาลใน batches จากถือถัง เช่น ตกขาวอาจจะเท่านั้นได้รับอนุญาตในกระแสน้ำสูง หากเป็นไปได้ รถถัง ควรผสม และในบางกรณีจำนวนการหมุนเส้น ถ้าดีผสมไม่ได้ มันอาจจะ จำเป็นเพื่อเพิ่มความถี่ของตัวอย่างในการปล่อย5.2 ส่วนประกอบหรือจุดตัวอย่างมีสองวิธีหลักเพื่อระบุสำหรับน้ำเสีย คอมโพสิต ( จุดตัวอย่าง :ตัวอย่างประกอบ สองประเภทของคอมโพสิตตัวอย่างที่ใช้กันทั่วไป , การไหลของสัดส่วนและเวลาตามสัดส่วน สำหรับการไหลแบบตัวอย่าง จำนวนตัวอย่างที่ถ่ายสำหรับแต่ละปริมาณของน้ำทิ้งที่กำหนดไว้ล่วงหน้า ( เช่นทุก 10 ลบ .) เวลาตัวอย่างตามสัดส่วน ,จำนวนตัวอย่างที่ถ่ายจากน้ำทิ้งของแต่ละหน่วยเวลา ( เช่นทุกm18 : การไหลของน้ำ และท่อระบายน้ำm18 รุ่น 4 พฤศจิกายน 2014 หน้า 48ชั่วโมง ) การวิเคราะห์ตัวอย่างประกอบทำให้มูลค่าเฉลี่ยของ determinand ในระหว่างระยะเวลาที่กลุ่มตัวอย่างได้รับการเก็บรวบรวม . มันเป็นเรื่องปกติที่จะเก็บคอมโพสิตตัวอย่างกว่า 24 ชั่วโมงเพื่อให้มูลค่าเฉลี่ยรายวัน เวลาที่สั้นกว่าสามารถใช้กับก่อนข้อตกลง จากเราการไหลแบบตัวอย่างที่ควรจะใช้เมื่อปริมาณน้ำที่ออกจากโรงพยาบาลแตกต่างกันอย่างมีนัยสำคัญทางสถิติทั้ง 2 ช่วง เมื่อปริมาณของตกขาวจะค่อนข้างคงที่แล้วเวลาคอมมีสัดส่วนที่เหมาะสมอุปกรณ์สุ่มอัตโนมัติแนะนำ ( ดูมาตรา 5.3 ) อย่างไรก็ตาม มันเป็นพิจารณาความมั่นคงของเป้าหมายรวมการเก็บตัวอย่างสาร ผ่าน เวลาตัวอย่างอาจเสื่อมขณะนั่งอยู่ในตัวอย่างอัตโนมัติอุปกรณ์ตัวอย่างจุด เหล่านี้คือตัวอย่างต่อเนื่อง ถ่ายที่สุ่มช่วงเวลาจากปลดและไม่ได้เกี่ยวข้องกับปริมาณของการปล่อย พวกเขาจะเหมาะที่สุดที่ :องค์ประกอบของน้ำเสียที่ค่อนข้างคงที่ปล่อยประกอบด้วยน้ำมันแร่ หรือสารระเหย หรือเมื่อเนื่องจากการสลายตัวการระเหยหรือการ เป้าหมาย สารจะไม่มั่นคงในตัวอย่างแยกขั้นตอนที่มีอยู่ ( ตัวอย่างเช่นน้ำมันชั้นลอยบนน้ำ )ต้องมีการตรวจสอบคุณภาพของน้ำในขณะนี้โดยเฉพาะ , ปกติการประเมินความสอดคล้องกับเงื่อนไขใบอนุญาตไหลไม่ต่อเนื่อง ( จากชุด หรือถือถัง ) แต่เมื่อน้ำคือผสมดีรวบรวมวัตถุขนาดใหญ่ และวัตถุที่ลอยไม่ใช่ตัวแทนของการอาจเป็นไปได้ยกคอมโพสิต เมื่อตัวอย่างขนาดใหญ่ขึ้นโดยคอมโพสิตชุดของตนเองตัวอย่างในห้องปฏิบัติการ มันเป็นสิ่งสำคัญที่จะต้องพิจารณาความมั่นคงของ determinand เป็นวัด ตัวอย่างเช่น , BOD จะเริ่มเสื่อมลงอย่างมาก หลังจาก 24 ชั่วโมงมันอาจจะไม่เหมาะสมที่จะรวบรวมตัวอย่าง รวมระยะเวลามากกว่า 24 ชั่วโมง เนื่องจากเพื่อความมั่นคงของ determinands แม้เมื่อ autosamplers เป็นห้องเย็น ตัวอย่างเช่นBOD , pH , COD และแอมโมเนีย5.3 อุปกรณ์และ mcerts ตัวอย่างอัตโนมัติเราต้องการให้อัตโนมัติ ตัวอย่างอุปกรณ์ที่ใช้สำหรับวัตถุประสงค์ที่ได้บรรยากาศการทดสอบและการรับรองเพื่อ mcerts perf
การแปล กรุณารอสักครู่..
