2.6. Analysis of sediment samples
Concentrations of the trace elements Cr, Cu, Zn, V and Ni were determined by
inductively coupled plasma optical emission spectrometry (ICP OES, PerkineElmer
Optima 2100 DV spectrometry, Massachusetts, USA) and Pb, Cd and As by
inductively coupled plasma mass spectrometry (ICP MS, Agilent Technologies
mod. 7500ce with Octapole Reaction System ORS, Santa Clara, USA) following
“aqua regia” digestion according to Bettinelli et al. (2000). Certified and experimental
values exhibited consistent values, recoveries being from 93 to 106% with
repeatability better than 8%. Standard reference materials were supplied by the
Community Bureau of Reference Sample (BCR): Certified Reference Materials CRM
277 and CRM 320 and 142 R. Mercury was determined by an automatic solid Hg
analyzer AMA 254.