ATTENTION
Inaccurate output power can also result from wear and tear of the equipment.
The problems due to wear and tear of the equipment can be avoided if the
equipment is undergoing periodic maintenance and testing by authorized
personnel.
ATTENTION
Muscular contractions are expected if HF surgery is done on structures capable
of electrical stimulation. (for example, during endoscopic surgery in urinary
bladder, near the obturator nerve.)
ATTENTION
This is applicable for split patient plate.
ATTENTION
The patient plate alarm is activated only if the neutral electrode is not properly
connected at the machine end or patient plate end, or if the percentage area of
contact of the patient plate is not sufficient.
ATTENTION
• Disposable handles are meant for single use only.
• Alcohol must never be used to clean and disinfect the front panel