Requirement:
ISO 13485:2003 Clause 8.4
The organization shall establish documented procedures to determine, collect and analyse appropriate data to demonstrate the suitability and effectiveness of the quality management system and to evaluate if improvement of the effectiveness of the quality management system can be made.
Description of Non-Conformity:
Data of management review input is not analyzed. Only raw data is presented.
Supporting Audit Evidence:
Management review input did not include the analysis of data. Only raw data is presented e.g. number of CARs found in each department, number of complaints received, etc.