6. Run Prime once more to verify the pressure is set correctly by observing that it stabilizes at the same pressure (+/-
0.1psi).
7. Switch the HTF back on.
8. Reset the Instrument Pressure Settings (RIPS) using one of the two methods:
• Customer utility as described in section 3.10.1 step 15
or
• LxMaintenance as described in section 3.10.2 step 13 (this method simpler and faster).