The extra-column variance results in a measured efficiency (N) that is approximately half of the actual column efficiency for a non- retained analyte, 6% less for a compound with a retention factor of approximately 2, and approximately 1% less for a compound with a retention factor >3.5. Therefore, the extra-column band broad- ening should be characterized in order to determine the actual performance of capillary LC columns.