5.5.1 The equipment shall have the capability of heating and
maintaining the temperature inside the bore of the cylinder in
accordance with the requirements specified in Table 2 throughout
the duration of the test.
5.5.2 The preferred method for calibrating the temperature
is to use a temperature sensor assembly having a sensor with at
least an accuracy of 60.08°C at 200°C and a 20 6 0.5-mm
long brass tip press fit on the end of the sensor. The diameter
of the brass tip shall closely match the diameter of the die and
the length of the active measuring length of the temperature
sensor (see Appendix X3).