6.6.2 The torque recording device shall be used to record the
signal from the torque measuring device and shall have a response time for full scale deflection of the torque scale of 1
s or less. The torque shall be recorded with an accuracy of 6
0.5 % of the range. Torque recording devices may include
analog chart recorders, printers, plotters, or computers.