1. Complaint No. : CR16-015
2. Product and Lot No. : Blood Tubing Set A105(P)/V671, Lot No.15L11, 15L17 and
15L18
3. Customer Name : Philippine Kidney Dialysis Center, Philippine
4. Content of Complaint : Some part of bloodlines glued together. After detached the tubing to mount to machine, they observed the tubing were either torn and dented.
5. Review of Device History Record (DHR)
We reviewed the documents of the concerned lots, there are no historical documentation that indicate the deviations.
6. Investigation of the retained samples
We conducted a visual inspection on the bloodlines from the reported lots (2 sets per lot). There was no sticking together of any parts and tubes.
7. Investigation of the complained samples
We received the complained samples 4 sets as following details.
• Blood Tubing Set A105(P)/V671, Lot No.15L11 : 2 sets
• Blood Tubing Set A105(P)/V671, Lot No.15L17 : 1 set
• Blood Tubing Set A105(P)/V671, Lot No.15L18 : 1 set
1. ร้องเรียนเลขที่: CR16-015
2 ผลิตภัณฑ์และ Lot No. : เลือดท่อชุด A105 (P) / V671, Lot No.15L11, 15L17 และ
15L18
3 ชื่อลูกค้า: ไตฟิลิปปินส์ศูนย์ฟอกเลือด, ฟิลิปปินส์
4 เนื้อหาของเรื่องร้องเรียน: บางส่วนของนํ้า glued ร่วมกัน หลังจากแฝดท่อต่อเชื่อมกับเครื่องพวกเขาสังเกตเห็นท่อมีทั้งฉีกขาดและเว้าแหว่ง. 5 ทบทวนประวัติศาสตร์อุปกรณ์บันทึก (DHR) เราได้ตรวจสอบเอกสารจำนวนมากที่เกี่ยวข้องไม่มีเอกสารทางประวัติศาสตร์ที่บ่งบอกถึงการเบี่ยงเบน. 6 การสืบสวนของตัวอย่างไว้ที่เราดำเนินการตรวจสอบภาพในสายเลือดจากจำนวนมากรายงาน (2 ชุดต่อจำนวนมาก) ไม่มีการติดกันของชิ้นส่วนใด ๆ และท่อ. 7 การสืบสวนของตัวอย่างบ่นเราได้รับตัวอย่างบ่น 4 ชุดมีรายละเอียดดังนี้. •เลือดท่อชุด A105 (P) / V671, Lot No.15L11: 2 ชุด•เลือดท่อชุด A105 (P) / V671, Lot No.15L17: 1 ชุด•เลือดท่อชุด A105 (P) / V671, Lot No.15L18: 1 ชุด