number of specimens that achieved the runout load, and the lowest failure load at which a specimen failed prior to reaching 5 million cycles. ASTM F2077 provides twodefinitions of specimen failure: (1) functional failure – permanent deformation rendering the IBFD unable to resist force, and (2) mechanical failure – onset of a newmaterial defect