NOTES:
1. ALL DIMENSIONS INDICATE 2D LOCATIONS OF
CONNECTORS, TERMINATIONS, AND/OR BREAKOUT
POSITIONS, NOT A DETERMINATION OF LENGTH
FOR INDIVIDUAL CABLES OR WIRES, WHICH
MUST BE DETERMINED BY THE SUPPLIER.
2. CONNECTORS TO BE ASSEMBLED AND TERMINATED
PER CONNECTOR MANUFACTURER SPECIFICATIONS.
3. CABLE/HARNESS TO BE LABELED PER GEHC
DOCUMENT 46-274914, CLASS V. LABEL PLACEMENT
AT APPROXIMATE CENTER OF CABLE.
4. MANUFACTURE AND TEST PER APPLICABLE
GE STANDARDS: 2269126GSP AND 46-297679.
5. VERIFY CENTER CONTACT SEATING DIMENSION
IS MAINTAINED WHILE APPLYING A MININUM
FORCE OF 10N (2.25lbf) AXIALLY AGAINST
THE CENTER CONTACT.
6. HI-POT TEST AT 4200 VAC FOR 3 SECONDS
BETWEEN CENTER CONTACT ON ITEM 1 (2x)
AND PLUG BODY OF CONNECTOR. MAXIMUM
LEAKAGE CURRENT PERMISSIBLE IS:
0,15mA FOR PI 5198312
0,20mA FOR PI 5198312-2
7. PART TO BE GEHC RoHS COMPLIANT PER
5240305GSP.
8. FOR BILL OF MATERIALS, APPROVED SOURCES,
AND LATEST REVISION OF PART(S) SEE GEHC
PRODUCT DATA MANAGEMENT SYSTEM.
NOTES:
1. ALL DIMENSIONS INDICATE 2D LOCATIONS OF
CONNECTORS, TERMINATIONS, AND/OR BREAKOUT
POSITIONS, NOT A DETERMINATION OF LENGTH
FOR INDIVIDUAL CABLES OR WIRES, WHICH
MUST BE DETERMINED BY THE SUPPLIER.
2. CONNECTORS TO BE ASSEMBLED AND TERMINATED
PER CONNECTOR MANUFACTURER SPECIFICATIONS.
3. CABLE/HARNESS TO BE LABELED PER GEHC
DOCUMENT 46-274914, CLASS V. LABEL PLACEMENT
AT APPROXIMATE CENTER OF CABLE.
4. MANUFACTURE AND TEST PER APPLICABLE
GE STANDARDS: 2269126GSP AND 46-297679.
5. VERIFY CENTER CONTACT SEATING DIMENSION
IS MAINTAINED WHILE APPLYING A MININUM
FORCE OF 10N (2.25lbf) AXIALLY AGAINST
THE CENTER CONTACT.
6. HI-POT TEST AT 4200 VAC FOR 3 SECONDS
BETWEEN CENTER CONTACT ON ITEM 1 (2x)
AND PLUG BODY OF CONNECTOR. MAXIMUM
LEAKAGE CURRENT PERMISSIBLE IS:
0,15mA FOR PI 5198312
0,20mA FOR PI 5198312-2
7. PART TO BE GEHC RoHS COMPLIANT PER
5240305GSP.
8. FOR BILL OF MATERIALS, APPROVED SOURCES,
AND LATEST REVISION OF PART(S) SEE GEHC
PRODUCT DATA MANAGEMENT SYSTEM.
การแปล กรุณารอสักครู่..