3.5. Caffeine RM homogeneityGo
Homogeneity of the prepared caffeine RM was assessed by studying the between- and the within-vial variability [[22], [23], [24], [25], [26], [27], [28], [29]]. Each of the randomly selected five vials was divided into three subsamples, each of which was measured three times by HPLC under repeatability conditions against the same calibration curve. The means of the measurement results were calculated, which are given in Table 3. The results were analyzed by analysis of variance (ANOVA) to estimate the uncertainty of the material variability and judge the material homogeneity. ANOVA results are given in Table 4, from which it can be seen that Fcalculated is smaller than Fcritical, which means that the RM is homogeneous. The uncertainty of the material homogeneity σh was calculated according to Eq. (1) and was found to be 0.0133, as reported in Table 3:
3.5. คาเฟอีน RM homogeneityGoHomogeneity ของคาเฟอีนพร้อม RM ได้รับการประเมิน โดยการศึกษาการระหว่าง- และความแปรปรวนภายในขวด [[22], [23], [24], [25], [26], [27], [28], [29]] แต่ละขวดห้าเลือกสุ่มถูกแบ่งออกเป็นสาม subsamples แต่ละที่โดยวัดสามครั้งจาก HPLC ภายใต้เงื่อนไขการทำซ้ำกับสอบเทียบโค้งเดียวกัน หมายของผลการวัดที่คำนวณ ซึ่งถูกกำหนดในตารางที่ 3 ผลลัพธ์ได้วิเคราะห์ โดยการวิเคราะห์ความแปรปรวน (ANOVA) การประเมินความไม่แน่นอนของความแปรปรวนที่วัสดุ และตัดสิน homogeneity วัสดุ ANOVA ผลถูกกำหนดในตาราง 4 ซึ่งจะเห็นได้ว่า Fcalculated มีขนาดเล็กกว่า Fcritical ซึ่งหมายความ ว่า RM เป็นเนื้อเดียวกัน ความไม่แน่นอนของ σh วัสดุเนื้อเดียวกันคำนวณตาม Eq. (1) และพบว่าเป็น 0.0133 ตามที่รายงานในตารางที่ 3: