T-262 Step Wedge Film and Densitometer
The densitometer shall be calibrated in accordance
with paragraph 6 of SE-1079, Calibration of Transmission
Densitometers, using a calibrated step wedge film
traceable to a national standard. The density of step
wedge comparison films and densitometer calibration
shall be verified by comparison with a calibrated step
wedge film.