contronExample: Correction of a temperature deviation in the load at 100°C.
1. Set calibration temperature CAL.2 to 100.0°C in SETUP and set the corresponding calibration
correction to 0.0°C.
2. Using a calibrated reference instrument, an actual temperature of 99.6°C is measured in normal
operation for a setpoint temperature of 100°C.
3. In SETUP set the calibration correction for CAL.2 to –0.4°C.
4. After the oven has settled down the reference instrument should read 100.0°C.
5. With CAL.1 a further calibration temperature can be programmed below CAL.2, and with CAL.3 an
additional calibration temperature above CAL.2.