Next, reconnect both high voltage leads to their
respective terminals inside the power supply and
disconnect them at the connection screw on cells door.
Support them away from any point of contact and
energize the power supply.
If either high voltage lead is defective the light will
indicate the trouble. Each lead may then be checked
separately by disconnecting them, one at a time, from
their respective terminals at the power supply. When a
lead is found to be defective, replace it in its entirety. Do
not repair or splice