Under a null weight condition, the output voltage of the load cell is not zero, and the output transduction ratio is not exactly at 4mV/kg. The tromer can be improved with providing an offset voltage to the input of the instrumental amplifier, and the latter can be improved with adjusting the R24 to increase the voltage gains.