The control of the immersion depth for the calibration is carried out by reducing the maximum possible immersion depth by 10 %.
The resulting heat dissipation error must not exceed 10 % of the measurement uncertainty aimed at, otherwise the measurement uncertainty must be increased, which usually requires a new, detailed uncertainty budget.