10.4.5 Corrective Action. If the BFB acceptance criteria are not met, the MS must be retuned. It may be necessary to clean the ion source, or quadrupoles, or take other necessary actions to achieve the acceptance criteria.
10.4.5 Corrective Action. If the BFB acceptance criteria are not met, the MS must be retuned. It may benecessary to clean the ion source, or quadrupoles, or take other necessary actions to achieve the acceptancecriteria.