5.2.6.10 Pressure gauges in the technical area of the sterilizer for utility lines readings: compressed air, inlet steam and water supplies.
5.2.6.11 Pressure switches to control utility lines: compressed air, inlet steam and water supplies.
5.2.6.12 Pressure gauges for sterile air filter housing reading.
5.2.6.13 Pressure switch to control steam pressure inside sterile air filter housing.
5.2.6.14 Jacket and chamber safety valves.
5.2.6.15 Thermostatic trap to remove jacket condensates.
5.2.6.16 All Critical Instruments must be calibrated and attached the reports.