To optimize the splitless hold time for a particular instrument, prepare a standard that contains both an early- and a late-eluting compound (e.g., fluorophenol and benzo(g,h,i)perylene). Inject this standard over a range of splitless hold times from 0.1 to 2.0 minutes and plot the data. An example of this optimization is shown in Figure 1.