The flow injection system developed for the research is
presented in Fig. 1. It consists of a multichannel peristaltic pump
(Gilson, France) equipped with Tygon pump tubes, a two-
positional eight-channel injection valve (Perkin Elmer, USA), a
glass mixing chamber of capacity 800 μL (made in our laboratory)
with a magnetic stirring bar, and an MR 1000 magnetic stirrer
(Heidolph, Germany). PTFE tubing (ID: 0.8 mm) was used as tubes
and loops. The signals were measured with the use of a model
SPEKOL 11 UV/Vis spectrometer (C. Zeiss, Germany) equipped with
a flow cell of length 10 mm (Zhaofa, China). The measurements
were performed in the transmittance mode. The operation of the
pump and valve was controlled by an electronic control device
(made in our laboratory). The dedicated software (developed in
our laboratory) served for data acquisition, real-time signal visua-
lization and peak parameters measuring
การไหลของระบบหัวฉีด เพื่อพัฒนาวิจัย
นำเสนอในรูปที่ 1 มันประกอบด้วย
ปั๊ม peristaltic หลายช่อง ( Gilson , ฝรั่งเศส ) อุปกรณ์ท่อปั๊ม tygon , สอง -
ตำแหน่งแปดช่องฉีดวาล์ว ( เพอร์กินเอลเมอร์ , USA ) ,
กระจกห้องผสม ความจุ 800 μ L ( ทำในห้องปฏิบัติการของเรา )
กับแม่เหล็กกวนบาร์ และนาย 1000 stirrer แม่เหล็ก
( heidolph , เยอรมัน )ท่อ PTFE ( id : 0.8 มม. ) มาใช้เป็นหลอด
และลูป สัญญาณจะถูกวัดด้วยการใช้แบบจำลอง
spekol 11 UV / VIS Spectrometer ( C . Zeiss เยอรมนี ) พร้อมกับ
ไหลเซลล์ยาว 10 มม. ( zhaofa , จีน ) การวัด
แสดงในโหมดการส่งผ่าน . การดำเนินงานของ
ปั๊มและวาล์วควบคุมการควบคุมอุปกรณ์อิเล็กทรอนิกส์
( ทำในห้องปฏิบัติการของเรา )ซอฟต์แวร์เฉพาะ ( พัฒนา
ห้องปฏิบัติการของเรา ) บริการข้อมูลเวลาจริงสัญญาณ visua -
รับรองเอกสารสูงสุด และค่าวัด
การแปล กรุณารอสักครู่..