Tocopherol (a-, c-, and d-tocopherols) isomers were separated
on a Supelcosil LC-Si column (250 4.6 mm, Supelco Inc., Supelco
Park, Bellefonte, USA). A 20-ll sample was injected with a mobile
phase consisting of a mixture of ethyl acetate/acetic acid/hexane
(1:1:198, v/v/v) at a flow rate of 1.5 ml/min. Detection was performed
at 295 nm using a UV detector. The unknown tocopherols
were identified on the basis of comparison of their retention
times with those of pure standards of a-, c-, and d-tocopherols.
The quantification was made on the basis of peak area of the unknowns
with those of pure standards of tocopherols (Sigma
Chemical Co.) using an external standard method. A D-2500 (Hitachi
Instruments Inc., Tokyo, Japan) Chromatointegrator model
with a built-in computer capability for data handling was used
for quantification purposes.
โทโคฟีรอ (a-, c- และ D-tocopherols) ไอโซเมอถูกแยกออก
ในคอลัมน์ Supelcosil LC-Si (250? 4.6 มม Supelco อิงค์ Supelco
พาร์คเบลลาฟอน, ประเทศสหรัฐอเมริกา) ตัวอย่าง 20 LL ถูกฉีดด้วยมือถือ
ขั้นตอนประกอบด้วยส่วนผสมของน้ำนม / กรดอะซิติก / เฮกเซน
(1: 1: 198, v / v / v) ที่อัตราการไหล 1.5 มิลลิลิตร / นาที การตรวจสอบการดำเนินการ
ที่ 295 นาโนเมตรโดยใช้เครื่องตรวจจับรังสียูวี tocopherols ที่ไม่รู้จัก
ถูกระบุบนพื้นฐานของการเปรียบเทียบของการเก็บรักษาของพวกเขา
ครั้งกับบรรดาของมาตรฐานความบริสุทธิ์ของ a-, c- และ D-tocopherols.
ปริมาณที่ถูกสร้างขึ้นบนพื้นฐานของพื้นที่จุดสูงสุดของราชวงศ์
กับผู้บริสุทธิ์ของมาตรฐานของ tocopherols (Sigma
เคมี จำกัด ) โดยใช้วิธีการมาตรฐานภายนอก D-2500 (Hitachi
Instruments Inc. , โตเกียว, ญี่ปุ่น) รุ่น Chromatointegrator
ที่มีความสามารถในตัวเครื่องคอมพิวเตอร์สำหรับการจัดการข้อมูลที่ถูกนำมาใช้
เพื่อวัตถุประสงค์ในเชิงปริมาณ
การแปล กรุณารอสักครู่..