4.1.1. Patient connection
1. The message "Please connect to the patient" that to connect to patient is displayed when dialyzer
filling is completed.
2. If turning in priming circulation mode: press the circulation button to stop the pump.
When "Clean-Treatment-Start" of "System-Setting-Prep./Rein." is "ON";
3. Pressing "OK" button in the message of to connect patient shows "Drain the solution in the blood line
with Clean-Treatment-Start? Yes/No".
4. When "Clean-Treatment-Start" function is not used, ordinary patient connect is performed. Connect
the arterial line to the patient, set the blood pump flow and select "No"'. Refer to post-No.5 procedure
"when "Clean-Treatment-Start" setting is "OFF" " for the further procedure.
5. Connect the arterial and venous lines to the patient and select “Yes” when "Clean-Treatment-Start"
function is used.
6. Blood pump and UF pump start operation with the set flows and blood removal starts.
7. Blood removal stops automatically once it reaches the set volume. It is also possible to perform the
blood removal additionally, following the guidance on the screen. Selecting “CTS Completed" exits the
blood removal screen.
8. Press the treatment button to start treatment.