The displacement of the crystal towards Line 3 seen in the scanning is confirmed by the X-rays. However, the scanning indicates an almost as large displacement orthogonally, towards Line 2, but this is not seen in the radiograph. There is no significant difference in count rate the measurements on the lateral surface, due to the displacement of the crystal. A difference in count rate is not expected, since the inside of the housing is at near vacuum and thus does not attenuate the photon beam, and the photon beam from the collimator is nearly parallel; an increase in distance will therefore not affect the fluence rate.