OPTIMIZATION OF HPLC METHOD
Optimization of mobile phase was performed based on
peak symmetry, peak width, and run time. The mobile
phase of water and acetonitrile (65:35 v/v) was found to
be satisfactory. The Fig.1 shows typical chromatograms
obtained from the analysis of a standard and sample
(fluconazole capsules, uncoated tablets and dispersible
tablets) solutions of fluconazole using the proposed
method. The retention time observed (2.47 min) permits a
rapid determination of the drug, which is important for
routine analysis. System suitability parameters for this
method are reported in Table1. The parameters were
within the acceptance limits