Table 2 shows that the use of the predictor electrode allows a more accurate measure of the sulfite concentration in the sample. Without compensation for these electroactive interferences, the channel biosensor over-predicts the concentration of sulfite in the sample, but when the corrected current from the predictor electrode is subtracted from the detector current, adequate recovery of spiked samples is obtained.